6
3.
INSTALLATION
The F6100 is supplied with a mounting kit which enables the instrument to be wall
mounted at the most suitable location to test your breathing-air / compressed-air
supply. The test location needs to be as close to the breathing-air supply point as
possible, in a well ventilated area free from dust and with a mains power supply
available. To install the bracket mark the hole locations on the wall and check with a
spirit level. Ensure you use the correct fixings to suit the wall.
Once you have installed the bracket mount the F6100 in place and connect the power
supply. It is important the unit is free from vibration during use.
4
BATTERY AND POWER SUPPLY
The F6100 is mains powered via an adaptor (220-240V, 9V output) which is supplied
with the unit, in case of a power failure an inbuilt rechargeable battery back-up system
will provide an additional 30 minutes operating time.
Note: Ensure the unit is switched off prior to connecting and disconnecting the mains
power supply.