Factair F6100 Operating Instructions Manual Download Page 20

19 

 

APPENDIX 2

 

 

Draeger Tube Tip Cutter 
 
Note: This item is not supplied as standard with the F6100. 
 

The  F2187  is  especially  designed  to  prevent  glass  from  falling  out  of  the  opener  by 
accident.  The reservoir for the broken-off tips is easy to empty.  

 
Instructions 

 
 
 
 
 
 
 
 
 
 

 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

1. 

Place the end of the tube 
between the 3 blades, and 
turn to score the end. 
 

 

2.  Push the tube at an angle 

to break the tip. 

 

 

3.  Repeat steps 1 and 

2 with the other end 
of the tube 

 

Summary of Contents for F6100

Page 1: ...0 F6100 SAFE AIR MONITOR OPERATING INSTRUCTIONS FAC QAM 097 Issue E 03 21...

Page 2: ...1...

Page 3: ...ART UP 7 5 2 F6100 SCREEN MENUS 8 5 3 USING THE TEST PORT 9 6 F6100 PC SOFTWARE 13 6 1 INSTALLING THE SOFTWARE 13 6 2 USING THE SOFTWARE 13 7 DATALOGGED RESULTS 17 APPENDICES Appendix 1 Helpful tips 1...

Page 4: ...providing the concentration in mg m3 It also includes digital airline pressure and ambient temperature readings Periodic oil measurements are completed using the Draeger Impactor which is inserted int...

Page 5: ...61 TUB0068 0 2 6 ppm Sulphur Dioxide SO2 6728491 TUB0014 1 25 ppm Polytest CH28401 TUB0017 Sensitivity dependant on compound present The tester can also be used with the old type Draeger reagent tube...

Page 6: ...D Training Video d USB Memory Stick with PC Power On Air Inlet Volt free Contact for remote alarm The current ratings are 0 5A at 125Vac and 1A at 24Vdc Mains input USB Connection for PC Software Wall...

Page 7: ...tions on the wall and check with a spirit level Ensure you use the correct fixings to suit the wall Once you have installed the bracket mount the F6100 in place and connect the power supply It is impo...

Page 8: ...s calibrated the instrument will then display the home screen At this point you can then connect the breathing air supply to the air inlet Please note that the tester can be piped to any pressure betw...

Page 9: ...for the configuration settings or to perform a Draeger Impactor or tube test press the arrow key in the top right hand corner The sub menu displays the date when the test port was last used and the r...

Page 10: ...Factair Part No Sensitivity Oil Impactor 8103530 TUB0053 0 1 mg m3 Oil chemical reagent tube 6728371 TUB0003 0 1 mg m3 Nitrous Fumes NOx 8103661 TUB0068 0 2 6 ppm Sulphur Dioxide SO2 6728491 TUB0014...

Page 11: ...sted Once the purge has finished the Draeger Impactor should be inserted into the test port and tightened firmly ensure the Impactor has an intact protective seal This seal must remain in place for th...

Page 12: ...from oil sulphur dioxide nitrous fumes or the polytest tube If you are intending to complete an oil tube test then you will need to input the test time If you can determine with a high degree of cert...

Page 13: ...o the test Ensure the arrow printed on the tube is pointing away from the instrument Press the arrow in the top right corner to proceed The test will proceed as described in section 3 and at the end o...

Page 14: ...alled when you run the application the following screen will be displayed Once installed on your computer run this program and then with the USB cable provided connect your PC or laptop to the instrum...

Page 15: ...yed to confirm you have connected the cable press ok to proceed A message will then be displayed to confirm the settings have been downloaded into the software click ok to proceed A typical example of...

Page 16: ...nternal Sensors In the Internal Sensors section you can select from a range of international breathing air standards The menu display the current test standard and you will be able to select a range o...

Page 17: ...ther assistance in configuring these sockets please contact Factair Updating the F6100 settings Once you have made all the required changes click on the Write to F6100 button A message will then be di...

Page 18: ...e website F6100 Data Monitoring Software http www factair co uk downloads The datalogging software makes it easy to analyse results identify trends and exceptions Results can be imported into the soft...

Page 19: ...strument off press the button in the top right hand corner of the screen The following warning will be displayed if there is insufficient pressure being supplied for the tube test The instrument requi...

Page 20: ...ally designed to prevent glass from falling out of the opener by accident The reservoir for the broken off tips is easy to empty Instructions 1 Place the end of the tube between the 3 blades and turn...

Page 21: ...and quickly identify minute quantities of oil For tests that pass with an oil concentration of less than 0 05mg m3 the screen will remain blank For tests above that read as follows The bottom line re...

Page 22: ...ll show a light brown or darker discoloration Synthetic oil The white crystals will show a yellow discoloration Note best seen by comparing with an unused tube 1 Using the Tube Tip Cutter place the am...

Page 23: ...ting the tube reading is multiplied by a factor of 10 i e a tube reading of 150mg m becomes 1500mg m To establish the pressure dewpoint refer to the graph on the back of the Safe Air Tester Result She...

Page 24: ...oxide The total length of the white section is the measure of concentration read directly from the 1 25ppm scale in parts per million Polytest CH 28401 REQUIREMENT NO DISCOLORATION The media in the de...

Page 25: ...ill or incineration OR 2 Place the tubes in a sharps or glass bin wearing protective gloves and safety spectacles Dispose of the bin via the company s normal hazardous waste disposal method s ie landf...

Page 26: ...25...

Page 27: ...26 Factair Ltd 49 Boss Hall Road Ipswich Suffolk IP1 5BN UK Tel Sales 44 0 1473 746400 Tel Hire 44 0 1473 746444 Fax 44 0 1473 747123 Email enquiries factair co uk www factair co uk...

Reviews: