3. PRECAUTIONS
3-1. Line voltage selection
This instrument must be operated with the correct line voltage selector switch setting
and the correct line fuse for the line voltage selected to prevent damage.
The instrument operates from either a 90V to 132V or 198V to 250V line source.
Before line voltage is applied to the instrument , make sure the line voltage selector
switch is set correctly.
In the case of the line voltage selector switch selected , must be operated to the
correct line fuse.
To change the line voltage selection
1. Make sure the instrument is disconnected from the power source.
2. Pull out the line voltage selector switch on the bottom panel.
Select the arrow mark position of the switch from Table 3-1.
Select the arrow mark to the desired position and plug it in.
3. Pull out the line fuse holder containing the fuse for overload protection.
Replace the fuse in the holder with the correct fuse from table3-1 & plug it in.
AC 90 ~ 110
AC 108~ 132
AC 198~ 242
AC 207~ 250
For continued protection against fire,replace the line fuse only with a fuse
of the specified type and rating.
3-2. INSTALLATION & HANDLING PRECAUTIONS
When placing the power supply in service at your workplace,observe the following
precautions for best instrument performance and longest service life.
1. Avoid placing this instrument in an extremely hot and cold place.
Specifically, don't leave this instrument in a close car,exposed to sunlight in
midsummer, or next to a space heater.
Table 3-1. Line voltage Selection & Fuse Ratings
T1.6A
[4]
GP-4303D/A , GP-4185D/A
T3.15A
230V
115V
Select Switch Mark
Fuse Ratings (250V)
Line Voltage
!!