                  6-3  PARALLEL  OPERATION

                            (1)  When  a  current  capacity  larger  than  a  single  unit  is  required,  a  couple  of  the  units  may  be
                                        used  in  a  parallel  connection.

        In  a  parallel  operation,a  unit  becomes  the  main  unit  and  other  becomes  the  sub-unit.
      Setting  of  the  voltage  and  current  should  be  made  of  the  main  unit.

                            (2)  Connection

    

Make  the  connection  while  the  power  off.

                            -                      GND                +

          -                  GND                +

                  (main)

                              (sub)

                            For  parallel  operatio,.make  the  connection  in  the  sequence  of  the  (+)  terminal  of  the  main  unit,(+)
                            terminal  of  the  sub-unit,(-)  terminal  of  the  main  unit,and(-)  terminal  of  the  sub-unit.
                          Then  connect  the  wires  between  the  Parallel  terminal  (main  unit)  and  the  Parallel  terminal  (sub  unit)
                          Use  thick  wires  for  all  the  connections.

                              (a)  Turn  the  voltage  knob  (5),(4)  and  the  current  knob(3)  of  the  sub  unit  full  clockwise.

                              (b)  The  voltage  and  current  are  variable  by  the  main  unit.
                                        The  output  current  limit  can  be  selected  within  a  range  of  about  10  to  100%  of  the
                                        double  of  the  maximum  rating  current.
                                        In  this  case,as  the  load  current  increase,  the  ampere  meters  of  both  main  and  sub  units
                                        swing,and  the  load  current  is  indicated  as  the  sum  of  both  amperemeters.

                            NOTES  :  Other  type  of  regulated  DC  Power  supply  can  be  connected  for  parallel  operation.

                      The  maximum  number  of  units  for  parallel  operation  is  limited  to  2  units.

      7.  NOTES  ON  ENVIRONMENTAL  CONDITION

            7-1.  Avoid  using  the  unit  in  such  a  place  where  the  ambient  temperature  exceeds  40℃   or  under
                          the  direct  sum  shines.            
                          Limit  the  maximum  output  current  when  the  unit  is  used  in  such  a  place  where  ventilation  is
                          interrupted  or  a  radiation  exists  other  equipments.

            7-2.  Use  the  instrument  whitin  10%  tolerance  of  the  specified  frrom  the  power  source.

Fig.6-4

[10]

LOAD

Summary of Contents for GP-4185A

Page 1: ... POWER SUPPLY GP 4303D 3 00 A 30 0 V CURRENT FINE VOLTAGE POWER GND GND C V C C DC POWER SUPPLY GP 4303A R POWER CURRENT FINE VOLTAGE 99 Washington Street Melrose MA 02176 Fax 781 665 0780 TestEquipmentDepot com ...

Page 2: ...e 0 70 Storage Humidity less than 85 2 For quality improvement the exterior design and specifications of the product can be changed without notice 3 Should any further information be required please contact the EZ DIGITAL company or sales outlet WARRANTY Warranty service covers one year the date of original purchase In case of technical failure a year repair service will be provided by EZ DIGITAL ...

Page 3: ...nd overload protection functions are built in to protect the over current The supply features anloge voltmeter and amperemeter enabling the user to monitor the supply s operation at a glance LEDs for Constant current C C and Constant Voltage C V enables monitoring of operation status Main input voltage AC115 or 230V selectable and 50 60Hz are supported 2 FEATURE Built in output current limit circu...

Page 4: ...ch from Table 3 1 Select the arrow mark to the desired position and plug it in 3 Pull out the line fuse holder containing the fuse for overload protection Replace the fuse in the holder with the correct fuse from table3 1 plug it in AC 90 110 AC 108 132 AC 198 242 AC 207 250 For continued protection against fire replace the line fuse only with a fuse of the specified type and rating 3 2 INSTALLATI...

Page 5: ...on near the instrument 10 Do not place this instrument face down on the ground or damage to the knobs may resuit 11 Do not connect other power source to of the output terminal 12 Ground terminal of power code must connect to the ground 13 According to output porarity it solved problem ESD and Floating voltage using the connect ground port to terminal of output 14 To connect the ground be careful a...

Page 6: ... Short circuit Bar Manual Fuse Power source AC 115 230V 50 60Hz Dimension 131 W x 150 H x 273 D Weight Approx 4 5Kg Protection circuit Ambient temperature 5 40 Operating Humidity 50 80 Insulation between Compensation 1 1digit 2 5 F S Ampermete Accuracy Voltmeter Accuracy 1 1digit 2 5 F S Ripple Voltage Less than 2mVp p Line Regulation Load Regulation Less than 0 01 3mV for load variation of 0 to 1...

Page 7: ...VR The fine adjustment knob of the output voltage 8 COARSE VR The coarse adjustment knob of the output voltage 9 OUTPUT TERMINAL The negative side of output terminal 10 OUTPUT TERMINAL GND When the positive polarity or negative polarity is to be connected to the ground either an appropriate output terminal should be connected to the GND by a short circuit bar supplied as accessory When no bar conn...

Page 8: ... and set the current while watching the amperemeter indication At this time set the voltage adjustment knob 5 full counter clockwise and set the white mark of the FINE adjustment knob 4 at middle The minimum current that can be set is as follows GP 4303D A Approx 200mA 3A GP 4185D A Approx 350mA 5A After setting a limiting current remove the short circuit wire Then set the voltage to a required vo...

Page 9: ...peration between other types is not available 2 Notes on series operation When the series connected two units of regulated DC power supplies are overloaded the unit in which the overcurrent protection circuit activated first is applied with the reverse voltage from the output of the unit and so the series control element of the former will be damaged To protect this problem connect two diodes betw...

Page 10: ... knob 3 of the sub unit full clockwise b The voltage and current are variable by the main unit The output current limit can be selected within a range of about 10 to 100 of the double of the maximum rating current In this case as the load current increase the ampere meters of both main and sub units swing and the load current is indicated as the sum of both amperemeters NOTES Other type of regulat...

Page 11: ...ut limit current caan be set in a range from 10 to 100 of the rating current and when the output current reaches the set value the instrument operates in the constant current condition As the output current comes down below the set value the constant voltage condition is automatically resumed 9 TO REPLACE THE FUSE Pull out F1 below with screw driver and get rid of F1 take out spare Fuse in the hol...

Reviews: