CXL-365 vario
corneal cross-linking
1025_IFU_507536 CXL-365 vario_EN_Rev 09
Page 31 of 49
All key pressings are confirmed by an acoustic signal, to see the proper reaction of the
device.
Document the measured number in the protocol and compare it with the setpoint. The setpoint
range is dependent on the energy setting. There are the following tolerances:
Adjustment of the Illumination
Tolerance range of the Illumination
3 mW/cm²
2,7 mW/cm
2
–
3,3 mW/cm
2
9 mW/cm²
8,1 mW/cm
2
–
9,9 mW/cm
2
18 mW/cm²
16,2 mW/cm
2
–
19,8 mW/cm
2
Figure 36: Setpoint range of the Illumination
In case the measurement is not in this interval it is not allowed to use the device for surgery. Please
contact your distributor.
4.6
Positioning of the device while application
Remove measurement adapter from the device (Figure 37). Push the adapter horizontally, until you
can remove it.
Figure 37: Remove the measurement adapter horizontally
Position the device at the horizontal bar and fix it (Figure 38).