4-4. Connection of Utility
(1) Check the model of product, voltage of mains
supply to be connected, phase, and ampacity.
The rating of mains supply to which the
product is connected is as shown on the right.
(2) Note that no power plug is attached
to the product.
Connect the power cord to the power board
directly. Otherwise, provide the cord with the
locking plug as shown in the figure on the right.
Be sure to ground the earth conductor securely
in either case.
When connecting the power cord to
the power source, do not use split socket
or power strip.
- 16 -
Connection should be made confirming
voltage, phase, capacity, type of outlet.
WARNING
Connect earthing conductor correctly.
WARNING
Do not use split socket or power strip.
WARNING
Product Model
FDU-1110
Power necessary for connection
Ampacity
10 A
Voltages
FDU-2110
AC220 V
Single phase
7 A
Power Cord
Recommended Plugs (Lock-type)
In case power plug is not attached
Wrong power connection may cause
fire or electrical shock.
To prevent electrical shock, never connect
the earthing conductor to the pipes for gas
or water.
Doing so may cause overcurrent,
and result in burnout of cables, or fire.
220 V AC
Single phase
(Catch Cap)
AC220 V
Single phase