c. Input the value which you want to calibrate in the Relative Humidity area
c-1. Hunit[%]
Value which product shows
c-2. Hstd[%] Standard value of calibration
1
Please enter calibration points in the blank, 10 points maximal.
2
The interval between two points should be 10% above.
d. click Relative Humidity > WRITE
c.
d.