THS30 Series Multifunction Temperature & Humidity Transmitter Operation Manual
PAGE
:
58 OF 65
Version
:
03
Date
:
2017.08.25
2. Put the product in the environmental of humidity which you want to calibrate
3. Wait the environmental of humidity is becoming stable
4. Retain the factory setting
a. Input the value which you want to calibrate in the Relative Humidity area
a-1. Hunit[%]
Value which product shows
a-2. Hstd[%] Standard value of calibration
1
Please enter calibration points in the blank, 10 points maximal.
2
The interval between two points should be 10% above.
3
Execute step 6 when you want to repeat the calibration points and factory setting points
or the humidity is less than 10%.
b. click Temperature > WRITE
b.