EYC FUMD Operation Manual Download Page 7

 

www.eyc-tech.com

 

 

 

6

 

 

 

Clamp-on Ultrasonic Flow 

 

Flow 

 

0.

 

Scale factor 
Refers to the ratio between 

actual value

 and 

reading value

. For 

example, when the measurement is 2.00, and it is indicated at 1.98 on the 
instrument, the scale factor constant is 1.01. 

1.

 

4 … 20 mA CAL:Check if the current loop has been calibrated 
before leaving the factory. Press

move

to display 4 mA 

or 24 mA, and at the same time, check with an ammeter to 
verify that Current Loop output displayed values. It is necessary to re-calibrate the 
current loop, if over the permitted tolerance. 

2.

 

Set zero Press

; reset 

Zero Poin

t”

 which was set by the 

user. 
 

3.

 

Low flow cut:Flow rate falls below the low flow cutoff value. 
 
 

The flow indication is driven to zero. This function can prevent the flow meter from 
reading flow after a pump as shut down but there is still liquid movement in the pipe, 
which will result in totalization error.   
Generally, 0.03 m/s is recommended to enter as the low flow cutoff point. The low flow 
cutoff value has no relation to the measurement results once the velocity increases over 
the low flow cutoff value.   

Setup Menu-Output 

Press

, Select 3. Output setting, and then

display: 

0.

 

RS-485 setup 
The window used to set serial port. It connection with the equipment of 
its serial port set of parameters must match. The first choice of data that 
baud rate:2400, 4800, 9600, 19200 choice. The second option that in check, None. 
Data length fixed to eight; Stop bit for a fixed length; Factory serial port parameters 
for the default 

9600, 8, None, 1

1.

 

4 ... 20 mA range 
Set the Current loop output value according to the flow 
value at 4 mA, and 20 mA. The flow unit is m

3

/h. 

2.

 

Alarm value (Option) 
Enter the low alarm value; any of the measured flow, which is lower 
than the low value, will activate the alarm in the OCT hardware or relay 
output signal. Enter the high alarm value; any of the measured flow, which is higher 
than the high value, will activate the alarm in the OCT hardware or relay output 
signal. 

Summary of Contents for FUMD

Page 1: ...Operation Manual eYc FUMD Clamp on Ultrasonic Flow Transmitter eYc XXXXX eYc FUMD...

Page 2: ...ponent P 02 3 Installation and Connection P 03 4 Panel Function P 03 5 Powering on P 04 6 Keypad Functions P 04 7 Window descriptions P 04 Display Menu P 04 Setup Menu P 04 Setup Menu Pipe parameter P...

Page 3: ...exert the best performance If you make a mistake there will be affected the transmitter s working and reduce the transmitter s life or cause some malfunctions Product component Inspection should be ma...

Page 4: ...Ultrasonic Flow Flow Installation and Connection Once the flow transmitter is installed it can be connected You can be found the cable have 5 pin terminals as per wiring diagram to connect 4 20 mA ou...

Page 5: ...s menu and scroll up and down to select the menu when press move to next digit press and the numbers scroll from 0 9 you can select the number Press to confirm Window descriptions Display Menu When th...

Page 6: ...select YES or NO After YES is selected 4 Time set Generally it is unnecessary to modify date time as the system is provided with a highly reliable perpetual calendar chip 5 System lock Once the system...

Page 7: ...error Generally 0 03 m s is recommended to enter as the low flow cutoff point The low flow cutoff value has no relation to the measurement results once the velocity increases over the low flow cutoff...

Page 8: ...h com 7 Clamp on Ultrasonic Flow Flow Tel 886 2 8221 2958 Web www eyc tech com e mail info eyc tech com Sustainable Green Professional Temp Humid Dew Point Air Velocity Volume Flow Pressure Measuring...

Reviews: