![EYC FTS34 Operation Manual Download Page 10](http://html1.mh-extra.com/html/eyc/fts34/fts34_operation-manual_2457474010.webp)
www.eyc-tech.com
9
Thermo Air Velocity Transmitter
Air Velocity
V.
Inspection and maintenance
1.
Maintenance
Since this product is inspected and calibrated for high accuracy at the factory before
shipment, no calibration on the installation site is necessary when this product is
installed. For inspection and maintenance follow the instructions below:
Periodic inspection
Periodically inspect this product for its sensing accuracy, and clean the cover. Set the
period between inspections based on atmospheric dust and other contaminants in the
installation environment.
2.
Troubleshooting
Sensor maintenance
Do not damage sensor surface during the maintenance process.
Troubleshooting
If any problem occurs during operation, refer to the table below for appropriate
solutions.
Problem
Cleck items
Soluations
No output
Unstable output
Disconnected wiring
Loose wiring
Power supply voltage
Sensor damages
Re-perform wiring
Crew on terminal tightly or replace wires
Replace the sensor
Slow response to
output
Errow in output
Moisture /condensation on the
product
Check installed location
Check installed angle
Check dust and contamination
on the sensor
Remove the sensor and filter. Dry
power-off state sensor in clean air
seasoning
Refer to the section
Align measurement head with flow
direction
Cleaning the filter
Changing the filter
Calibrate
Replace the sensor