5.2

 

Transmitter-Module 

All gas sensing equipment is combined in one exchangeable transmitter-module which forms 

the right-hand part of the wall mounted housing. In case of maintenance this module can easily 

be taken off from the Bio-Compact and replaced by a new pre-adjusted module.  
There are two types of transmitter modules available: 
 

Transmitter module 

Article 

(new/replacement) 

In combination with: 

CH

4

, O

2

, H

2

S (H) 

Measuring ranges:  
0-100 % (v/v) Methane 

0-25 % (v/v) Oxygen  
0-1000 ppm Hydrogen Sulphide 

290010/ 

295010 

Bio-Compact 

Type (H) 

CH

4

, O

2

, H

2

S (L) 

Measuring ranges:  
0-100 % (v/v) Methane 

0-25 % (v/v) Oxygen 
0-100 ppm Hydrogen Sulphide

 

290011/ 

295011 

Bio-Compact 

Type (L) 

 

5.2.1

 

Replacement of a Transmitter Module 

On delivery of the Bio-Compact a new transmitter module is installed (Art. 290xxx). In case 

you need a new one you will get a replacement transmitter module (Art. 295xxx), on which all 

wear parts have been replaced. The module has completely been checked and adjusted. It has 

the  same  warranty  as  a  new  module.  In  case  you  return  the  replaced  module  you  will  get  a  

credit note for it. 
Replacement is done in few steps: 
1.

 

Release access level 3 on the control unit (see instruction manual of the control unit). 

2.

 

Press the keys "F1+F5+F6". On the display "CHANGE MODULE" is indicated. 

3.

 

Select "Change". The latest actual measuring value is frozen and is constantly sent to the 

connected periphery for the duration of the module exchange.  

4.

 

Replace Module:  

a.

 

Untighten glands at measured gas inlet and outlet and remove hoses. 

b.

 

Open the housings of control unit and transmitter module. 

c.

 

Disconnect both cable plugs placed at the control unit.  

d.

 

Loosen the four fixing screws of the transmitter module. 

e.

 

Remove transmitter module. Be careful not to damage the two ribbon-cables which 

connect the transmitter module to the control-unit. 

f.

 

Install new transmitter module in reverse order.  

5.

 

Select "Ready". Measurement is started again. 

6.

 

Select "End". The Bio-Compact returns to the standard indication. 

The Transmitter Module is pre-adjusted. After exchange it is necessary to let it run for approx-

imately 1 to 3 hours till the final measuring accuracy is reached. During this time you should 

not perform any adjustment. 

Summary of Contents for Bio-Compact

Page 1: ...Instruction Manual ExTox Biogas Analyser Bio Compact...

Page 2: ...Page 2 of 12 BA_Bio Compact_e_2015 10 22 doc...

Page 3: ...h reliability due to modern sensor techniques and consequent interpretation acc to the requirements of the regulations and standards valid for industrial application The Bio Compact combines all neces...

Page 4: ...ays and Digital Inputs 8 5 1 3 Calibration and Adjustment 8 5 2 Transmitter Module 10 5 2 1 Replacement of a Transmitter Module 10 6 Operation of Bio Compact 11 6 1 Flow rate monitoring 11 6 2 Flame A...

Page 5: ...formance of work when mainte nance and repairs are done by them On receipt of goods please take care that packing and consignment are not damaged and the goods supplied correspond to the articles desc...

Page 6: ...Detection Systems Furthermore they are in conformity with the Euro pean Directives 94 9 EC ATEX 2004 108 EEC EMC and 2006 95 EEC LVD 2 3 Indications and Facilities The control unit ET 4D2 covering th...

Page 7: ...t changes to the phase flushing and starts the measurement later on For channels in TIMER Mode all measured values alarms and outputs remain frozen on the latest value of the previous measuring phase...

Page 8: ...Channel Configuration In the menu the mode TIMER is factory sided activated Attention Changing of these setting should in no way be done by the user Changed measur ing functions may otherwise lead to...

Page 9: ...resulting from this falsifies calibration adjustment 4 READ SENS The measured gas pump is switched off and batch meas urement is done within two minutes 5 READY Calibration is completed and a plausibi...

Page 10: ...justed It has the same warranty as a new module In case you return the replaced module you will get a credit note for it Replacement is done in few steps 1 Release access level 3 on the control unit s...

Page 11: ...rotected against weather conditions and bigger climatic fluctua tions In general we do not recommend an installation outside as in principle measuring accuracy and lifetime are influenced in a negativ...

Page 12: ...ervoltage All connections are inside the housing of the control unit ET 4D2 accessible via the cable glands at the bottom of the housing Connection of transmitters and additional devices gas suction p...

Reviews: