Extech Instruments SL130 User Manual Download Page 5

Version 3.1   1/06 

Range Selection 

Slide the Range switch to the 30-80, 60-110 or the 80-130 position. 
The selected range will be indicated on the LCD bargraph. If the 
measured dB level exceeds the range selected “OL” will be 
displayed. If the measured dB level is below the selected range 

“- - -

“ will be 

displayed. 

Normally, select the range in which the upper dB level is greater than any 
expected sound pressure level. The 60 to 110dB range is the most commonly 
used range. Use the 30 to 80dB range in quieter areas such as offices and 
classrooms. 

Note: When the range is changed the Max display may indicate an overload “OL”. Press 
the MAX RESET button to clear the MAX display. 

High Limit Set 

Press the up or down LIMIT arrow buttons to set the High/Low limit 
set=point as displayed on the LCD. If the measured dB level 
exceeds the set limit, the          indication will appear in the display 
and the over limit signal will appear at the alarm output connector. 
The display indication and output signal will remain active as long as the sound pressure 
level exceeds the set limit. 

High/Low LED Alert 

The four LEDs are used to alert users when a sound level either exceeds or goes below 
the Limit set-point. This feature is available only while the AC adaptor is used to power the 
meter. It is disabled under battery power. 
The LEDs will provide a steady or flashing colored indication depending on the setting of 
the side panel ALARM and Alert switch settings. 

The three Alert switch positions are: (1) RED, (2) RED-GREEN, (3) GREEN 

 
 
 
 
 

Alarm 
Switch 

Alert Switch  

LED response 

Red (1) 

LED’s flash red while dB level exceeds set point. 

Red-Green (2) 

LED’s flash red while dB level exceeds set point.  
40 second green indication when level transitions from 
high to low. 

High 
ALARM

 

Green (3) 

40 second green indication when level transitions from 
high to low. 

Green (3) 

LED’s are green while dB level is below set point. 

Red-Green (2) 

LED’s are green while dB level is below set point.  
40 second flashing red indication when level 
transitions from low to high. 

Low 
ALARM

 

Red (3) 

40 second flashing red indication when level 
transitions from low to high. 

 

 

Summary of Contents for SL130

Page 1: ... Type 2 Sound Level Meter Standards Settable High Limit with large bright High Low indication and Alarm Output Wall or Desk mountable Extra large LCD display with bargraph Maximum level display with time stamp Optional alarm kit drives remote over limit indicators ...

Page 2: ...OFF time extends battery life to 30 days when programmed 8 hours per day The bright LED s are functional only when the AC power adaptor is used Careful use of this meter will provide years of reliable service Specifications Applicable standards Meets IEC 60651 1979 and ANSI S1 4 1983 Type 2 SLM standards Display 11 7 x 7 94cm 4 6 x 3 125 multifunction LCD Frequency bandwidth 31 5 Hz to 8 KHz Micro...

Page 3: ...utton Rear Control Panel 1 Range select switch 2 Time set buttons 3 Limit set buttons 4 Calibration adjustment 5 A C select button 6 F S select button Side Panel 1 Alert switch Green position 2 Alert switch Red Green position 3 Alert switch Red position 4 High Low Alarm switch 5 Remote Alarm output connector 6 AC adaptor connector Rear Panel 1 Tilt stand 2 Battery compartment 3 AC Adaptor connecto...

Page 4: ...just Time of day Hour Time of day Minutes Time of day AM PM 24 Hour Year Month Day Auto ON Hour Auto ON Minutes Auto OFF Hour Auto OFF Minute A C Frequency Weighting Change the Frequency Weighting by pressing the A C button The A or C icon will appear on the display With A weighting selected the frequency response of the meter is similar to the response of the human ear A weighting is commonly use...

Page 5: ...play indication and output signal will remain active as long as the sound pressure level exceeds the set limit High Low LED Alert The four LEDs are used to alert users when a sound level either exceeds or goes below the Limit set point This feature is available only while the AC adaptor is used to power the meter It is disabled under battery power The LEDs will provide a steady or flashing colored...

Page 6: ... the power switch is set to the AUTO position Calibration The sound level meter should be calibrated on a regular basis to assure that the meter s performance and microphone sensitivity has not change Some operational directives require daily calibration checks OSHA Extech offers several acoustic calibrators to perform the calibration 1 Set the meter to the 60 to 110dB or 80 to 130dB range 2 Place...

Page 7: ...tension cable can then be plugged into the matching connectors Calibration should be performed after the cable is installed Alarm Output optional The alarm output drives the optional Remote Alarm Relay Module cable and relay when the LIMIT set point is exceeded The relay module can be used to power external warning signs QUIET signs or other indicating devices when the sound level is higher than t...

Page 8: ...t to be free of defects in parts and workmanship for one year from date of shipment a six month limited warranty applies on sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 ext 210 for authorization or visit our website at www extech com click on Contact Extech and go t...

Reviews: