SDL720-en-GB_V1.0   12/16

 

Getting

 

Started

 

Power

 

ON

OFF

 

 

Power

 

the

 

meter

 

by

 

pressing

 

and

 

holding

 

the

 

power

 

button

    

for

 

at

 

least

 

1.5

 

seconds.

 

 

Press

 

and

 

hold

 

the

 

power

 

button

 

for

 

at

 

least

 

1.5

 

seconds

 

to

 

power

 

OFF

 

the

 

meter.

 

 

This

 

meter

 

is

 

powered

 

by

 

six

 

(6)

 

1.5VDC

 

‘AA’

 

batteries

 

or

 

by

 

optional

 

AC

 

adaptor.

 

If

 

the

 

meter

 

will

 

not

 

switch

 

ON

 

please

 

check

 

that

 

fresh

 

batteries

 

are

 

installed

 

in

 

the

 

rear

 

battery

 

compartment

 

or,

 

in

 

the

 

case

 

of

 

the

 

AC

 

adaptor,

 

check

 

that

 

the

 

adaptor

 

is

 

connected

 

correctly

 

to

 

the

 

meter

 

and

 

to

 

an

 

AC

 

source.

 

Display

 

Backlight

 

To

 

turn

 

the

 

display

 

backlight

 

ON

 

or

 

OFF,

 

press

 

and

 

hold

 

the

 

backlight

 

 

button

 

for

 

at

 

least

 

1.5

 

seconds.

 

The

 

meter

 

will

 

beep

 

when

 

switching

 

the

 

backlight

 

ON

 

or

 

OFF

 

unless

 

the

 

beeper

 

is

 

disabled.

 

Units

 

of

 

Measure

 

The

 

currently

 

selected

 

unit

 

of

 

measure

 

is

 

shown

 

below

 

the

 

measurement

 

value

 

on

 

the

 

meter’s

 

LCD.

 

To

 

change

 

the

 

unit

 

of

 

measure,

 

press

 

and

 

hold

 

the

 

UNIT

 

button

 

until

 

the

 

desired

 

unit

 

of

 

measure

 

appears

 

and

 

then

 

release

 

the

 

UNIT

 

button.

 

The

 

meter

 

begins

 

scrolling

 

through

 

the

 

available

 

units

 

of

 

measure

 

(see

 

table

 

below)

 

after

 

the

 

UNIT

 

button

 

has

 

been

 

depressed

 

for

 

at

 

least

 

1.5

 

seconds.

 

 

UNIT

 

DISPLAY

 

INDICATOR

 

psi

 

PSI

 

Inch

 

Hg

 

In

 

Hg

 

Inch

 

H

2

O

 

In

 

H2O

 

hPA

 

hPA

 

kPA

 

_PA

 

Bar

 

bAr

 

Kg/cm

2

 

_g

 

C2

 

mm

 

Hg

 

‐‐

Hg

 

Meters

 

H

2

O

 

t

 

H2O

 

Atmospheres

 

AtP

 

ZERO

 

Adjustment

 

To

 

null,

 

or

 

zero,

 

the

 

display

 

for

 

a

 

connected

 

sensor,

 

press

 

and

 

hold

 

the

 

up

 

and

 

down

 

arrow

 

keys

 

for

 

at

 

least

 

1.5

 

seconds.

 

The

 

meter

 

will

 

show

 

CAL

 

0

 

on

 

the

 

bottom

 

left

 

of

 

the

 

screen.

 

Data

 

Hold

 

To

 

freeze

 

a

 

displayed

 

reading

 

on

 

the

 

LCD,

 

momentarily

 

press

 

the

 

HOLD

 

button

 

(the

 

HOLD

 

icon

 

will

 

appear

 

above

 

the

 

reading).

 

To

 

exit

 

HOLD,

 

press

 

the

 

HOLD

 

button

 

again.

 

www.calcert.com

[email protected]

1.888.610.7664

0

5

10

15

20

25

30

Summary of Contents for SDL720

Page 1: ...User Manual Heavy Duty Differential Pressure Manometer Model SDL720 Additional User Manual Translations available at www extech com www calcert com sales calcert co 888 610 7664 0 5 10 15 20 25 30 ...

Page 2: ...ustomer Support Safety International Safety Symbols This symbol adjacent to another symbol or terminal indicates the user must refer to the manual for further information Meter Description 1 P1 Input positive 2 P2 Input negative 3 LCD Display 4 HOLD and Backlight key 5 MAX MIN key 6 SET and Clock key 7 RS232 output jack 8 Reset button 9 Power Adaptor jack 10 SD card slot 11 ENTER and LOG key 12 Do...

Page 3: ...he currently selected unit of measure is shown below the measurement value on the meter s LCD To change the unit of measure press and hold the UNIT button until the desired unit of measure appears and then release the UNIT button The meter begins scrolling through the available units of measure see table below after the UNIT button has been depressed for at least 1 5 seconds UNIT DISPLAY INDICATOR...

Page 4: ... rate Hours Minutes Seconds PoFF Automatic power off management Enable or disable the auto power off function bEEP Set the beeper sound ON OFF dEC Set the numerical format USA decimal 20 00 or European comma 20 00 SdF Format the SD memory card Setting the Clock Time 1 Access the dAtE parameter 2 Use the arrow keys to change a value 3 Use the ENTER button to step through the selections 4 Press and ...

Page 5: ...ter 2 Use the arrow buttons to select USA or EUro 3 Press ENTER to confirm setting 4 Press and hold the SET button for at least 1 5 seconds to exit to the normal operation mode or simply wait 7 seconds for the meter to automatically switch to the normal operating mode SD Card FORMATTING 1 Access the Sd F parameter 2 Use the arrow buttons to select YES to format the card select NO to abort Note tha...

Page 6: ...the meter displays a positive pressure reading if the P1 pressure is greater than P2 and a negative reading if P2 is greater than P1 Datalogging Types of Data Recording Manual Datalogging Manually log up to 99 readings onto an SD card via push button press Automatic Datalogging Automatically log data onto an SD memory card where the number of data points is virtually limited only by the card size ...

Page 7: ...ero 2 Press and hold the LOG button for at least 1 5 seconds The meter will flash the DATALOGGER icon at the selected sampling rate indicating that readings are now being automatically recorded to the SD card 3 If a card is not inserted or if the card is defective the meter will display EMPTY and exit the DATALOGGER mode In this case switch the meter OFF and try again with a valid SD card 4 Pause ...

Page 8: ...to a PC SD card reader If the PC does not have an SD card slot use an SD card adaptor available at most outlets where computer accessories are sold 4 Power the PC and run a spreadsheet software program Open the saved documents in the spreadsheet software program see example spreadsheet data screen below Spreadsheet data example AC Power Adaptor This meter is normally powered by six 6 1 5V AA batte...

Page 9: ...se Couplings This meter is supplied with a pair of hose couplings They fit 4 0mm 0 157 tubing See diagram on how to connect these couplings to the hose and the meter www calcert com sales calcert co 888 610 7664 0 5 10 15 20 25 30 ...

Page 10: ...ery compartment cover with the two 2 Phillips screws Never dispose of used batteries or rechargeable batteries in household waste As consumers users are legally required to take used batteries to appropriate collection sites the retail store where the batteries were purchased or wherever batteries are sold Disposal Do not dispose of this instrument in household waste The user is obligated to take ...

Page 11: ...en logging at a 1 second rate MANUAL LOGGING Set the sampling rate to 0 seconds Select 1 to 99 locations Data error number 0 1 number of total saved data Memory Card SD memory card 1G to 16GB size Data Hold Freeze the displayed reading Memory Recall Record and Recall the Maximum and Minimum readings Display update rate Approx 1 second Operating Temperature 0 to 50 C 32 to 122 F Operating Humidity ...

Page 12: ...s H2O 0 5 20 40 m H2O 0 01 1 974 ATP 0 001 Conversions Useful conversion Factors From To Multiplier in of H2O in of Hg 0 07355 in of H2O cm of H2O 2 54 mm of Hg in of H2O 0 53524 Copyright 2016 FLIR Systems Inc All rights reserved including the right of reproduction in whole or in part in any form ISO 9001 Certified www extech com www calcert com sales calcert co 888 610 7664 0 5 10 15 20 25 30 ...

Reviews: