Summary of Contents for RH550

Page 1: ...USER MANUAL Humidity Temperature Chart Recorder with Touch Screen and Dew Point calculation MODEL RH550 ...

Page 2: ... up the RH550 5 3 2 Meter Power 5 3 3 Main Interface Screen 6 3 4 Main Operating Modes LIVE and RECORD 7 4 MAIN MENU 8 4 1 GENERAL SETUP 8 4 2 RECORDING SETUP 8 4 3 VIEW ALARMS 9 4 4 HISTORY 9 4 5 DISPLAY SETUP 10 4 6 DATE TIME SETTINGS 11 4 7 INFORMATION SCREEN 11 5 PC INTERFACE 12 6 SCREEN SHOT STORAGE 12 7 FIRMWARE UPGRADE 12 8 SPECIFICATIONS 13 ...

Page 3: ... 6 5 extension cable High resolution color touch screen LCD 1024 x 600 pixels with brightness control screen saver and sleep mode Real time monitoring live mode and datalogging record mode Screen Saver continues to show digital and graphic representation of real time readings Intuitive menu for setting alarms changing temperature units setting date time updating firmware configuring the datalogger...

Page 4: ...k for powering the meter and connecting to a PC 3 Temperature Relative Humidity sensor shown connected to the cradle jack 4 Power ON OFF button 5 Sensor extension cable RH550 C 6 Wall mount bracket 7 Desktop tilt stand 8 Sensor cradle jack shown with sensor installed 9 Wall mount screw holes 10 Wall mount bracket hinge 4 3 1 2 ...

Page 5: ...all first and then snap the RH550 onto the bracket Note that the wall mount bracket has a swivel arm 10 for convenient angling Note Mount the bracket with the hinge to the right in order for the display to be upright 3 2 Meter Power A rechargeable lithium battery installed at the factory powers the RH550 To switch the RH550 ON or OFF long press the power button 4 To recharge the battery plug the s...

Page 6: ...mode or RECORDING mode 8 Tap to Start or Stop Recording 9 Current cursor position tap on curves to move cursor position 10 Min Max values for each measurement parameter 11 Digital representation of readings 12 X axis for date and time cursor position 13 Y axis for measurement scale slide up and down to view entire scale 14 Color coded Temperature Humidity and Dew Point Temperature values tap anywh...

Page 7: ...ections for additional details Use the MAIN MENU Recording Options sub menu to set the sampling rate HH MM SS from 00 00 01 09 59 59 schedule a start time and set a limit on the number of data points to record 500 1k 5k 10k 20k or MAX Note that when you select MAX the RH550 creates multiple 20 000 data point files not one large file See the MAIN MENU section for details Once programmed tap the STA...

Page 8: ...e circle next to oF or oC Set the Alarm ON or OFF by tapping ON or OFF To set the HIGH and LOW temperature Alarm thresholds open the numerical keypad by tapping the small arrow to the right of each programming field Set the HIGH and LOW relative humidity Alarm thresholds in the same way as temperature Alarm thresholds described above Click SAVE to store your settings or click X to cancel and exit ...

Page 9: ...hat when you select MAX the RH550 creates multiple 20 000 data point files not one large file Click SAVE to store your settings or click X to cancel and exit the sub menu Note Non Recording mode Live has a fixed sample rate 1 minute 4 3 VIEW ALARMS Scroll through the history of Alarm events if any by swiping with one finger Sort the Alarm log by date by tapping the programming fields next to Sort ...

Page 10: ...ub menu 4 5 DISPLAY SETUP Set the screen brightness by dragging the brightness slider Set the screen saver ON OFF by tapping ON or OFF Set the screen saver timer by tapping the small arrow on the programming field to open the interactive numeric keypad Set the display Sleep Mode by tapping ON or OFF Set the Sleep Mode timer by tapping the small arrow on the programming field to open the interactiv...

Page 11: ...y tapping the small arrow on the programming field to open the interactive calendar 4 7 INFORMATION SCREEN View manufacturer name model number current firmware version date time of last firmware update See the Firmware Upgrade section for details on updating the RH550 firmware ...

Page 12: ...a folder will open Click on Android File Transfer folder and you will see the 4 folders from the RH550 DCIM Download Pictures RH T Data 6 Screen Shot Storage Swipe downward on the RH550 screen with three 3 fingers to store a screen shot of the current display If taking a screen shot of a MENU window place your fingers at the top of the Menu window and not at the top of the screen and then swipe do...

Page 13: ...ensor probe on extension cable 10 to 40 C 14 to 104 F 0 5 C 0 9 F Sensor on Tablet Dew point temperature range 20 to 80 C 4 to 176 F Dew Point temperature is a calculated value Internal storage 16GB Sample Rate Live mode Fixed at 1 minute Recording mode adjustable from 1 second to 23 hrs 59min 59sec AC Adaptor 5V 2A DC with universal adaptors Battery 3000mAh Lithium polymer battery Operation tempe...

Page 14: ...most of our products Contact us for information on calibration and repair availability refer to the contact information below Annual calibrations should be performed to verify meter performance and accuracy Product specifications are subject to change without notice Please visit our website for the most up to date product information www extech com Contact Customer Support Customer Support Telepho...

Reviews: