Set the Date and Time for a scheduled recording by tapping the small arrow on the
programming field. This will open an interactive calendar. Note that the default start time
is the current system date and time.
Select the maximum number of data points by clicking the small arrow on the
programming field. Select 500, 1k, 5k, 10k, 20k or MAX. If you select MAX, recording will
continue until you manually stop (by pressing STOP RECORDING on the Main Interface
Screen) or until it fills the internal memory (16GB). Note that when you select MAX, the
RH550 creates multiple 20,000 data point files, not one large file.
Click SAVE to store your settings or click X to cancel and exit the sub-menu.
Note: Non-Recording mode (Live) has a fixed sample rate (1 minute).
4.3 VIEW ALARMS
Scroll through the history of Alarm events (if any) by swiping with one finger.
Sort the Alarm log by date by tapping the programming fields next to ‘Sort by date’ to
open the interactive calendar.
Sort the Alarm log by high/low measurement values by tapping on the programming
fields next to ‘Sort by value’ to open an interactive keypad.
Delete alarm entries be tapping the entry and choosing Delete.
4.4 HISTORY
Use one finger to swipe (scroll) through the list of recorded datalog files.