SET BUTTON

 

The SET button is used to manually step through the 5 stored output values. 

1.

Select the SOURCE mode

2.

Press the SET button and the value stored in memory location 01 will be sourced. “MEM.01”
appears in the display.

3.

Each press of the SET button will step through the 5 memory locations.

4.

The arrow buttons can be used to adjust the value in each memory location.

STEP/MEM BUTTON

 

The STEP/MEM button is used to automatically step through the 5 stored output values. The meter 
can be set for a single cycle of the stored values or a continuous cycle. 

1.

Select the SOURCE mode

2.

Press and HOLD the STEP/MEM button. ”STEPSS” (single cycle) and “STEPSC”
(continuous cycle) will alternately appear in the display. Release the button when the
desired mode is displayed.

3.

In single cycle mode the meter will source the current displayed in MEM01 for 5 seconds.
The meter will then advance to MEM02 for 5 seconds. This will continue through MEM05
and then step down through the memory locations. The cycle will end when MEM01 has
been reached.

4.

In the continuous mode the cycle will continue until manually stopped.

5.

Momentarily press the STEP/MEM button to stop the cycle. “END” will briefly appear in the
display.

STORING VALUES INTO MEMORY

 

The default values stored in the memory locations are: 

Memory 

Location 

mA

%

M1 

4.00mA

0.0%

M2 

8.00mA

25%

M3 

12.00mA

50%

M4 

16.00mA

75%

M5 

20.00mA

100%

To change the values in memory: 
1.

Select the SOURCE mode

2.

Press the SET button to select the memory location to be changed.

3.

Press the arrow buttons to adjust to the new value

4.

Momentarily press the MEM button to store the value. The memory location icon will blink
while the value is being stored.

actoolsupply.com

actoolsupply.com

Summary of Contents for PRC10

Page 1: ...l PRC10 Extech PRC10 Current Calibrator Meter Extech PRC15 Current and Voltage Calibrator Meter Extech PRC20 Multi Thermocouple Calibrator Extech PRC30 Multifunction Process Calibrator actoolsupply com actoolsupply com actoolsupply com ...

Page 2: ...nput range Turn the unit OFF when the device is not in use Remove the batteries if the device is to be stored for longer than 60 days Never dispose of batteries in a fire Batteries may explode or leak Never mix battery types Always install new batteries of the same type Cautions Improper use of this meter can cause damage shock injury or death Read and understand this user manual before operating ...

Page 3: ...memory STEP button 9 Backlight ZERO button 10 Test lead inputs Display Layout 1 SOURCE mode icon 2 ZERO function status icon 3 MEASURE mode icon 4 Loop Power icon 5 Auto Power OFF active icon 6 Battery status icon 7 Measure mode value 8 Measure mode units icon 9 Datalogger memory location 10 Source mode value 11 Source mode units icon 3 4 5 6 7 8 10 9 2 1 actoolsupply com actoolsupply com ...

Page 4: ...TTON Press the UNIT button to select mA or units I O BUTTON Momentarily press the I O button to select either SOURCE output or MEASURE input MODE BUTTON Loop Power In the MEASURE mode Press and Hold the MODE I O button for 1 seconds to turn loop power on or off Backlight BUTTON Momentarily press the Backlight button to turn the backlight on or off ZERO BUTTON In the MEASURE or SOURCE mode Press an...

Page 5: ...ired mode is displayed 3 In single cycle mode the meter will source the current displayed in MEM01 for 5 seconds The meter will then advance to MEM02 for 5 seconds This will continue through MEM05 and then step down through the memory locations The cycle will end when MEM01 has been reached 4 In the continuous mode the cycle will continue until manually stopped 5 Momentarily press the STEP MEM but...

Page 6: ...ut Mode of Operation In this mode the unit can source current up to 24mADC at 1000 ohms The current can be output either manually or stepped from memory as described earlier 1 Turn the meter ON 2 Press the I O button to select SOURCE 3 Use the UNIT button to select or mA 4 Connect the Calibration Cable to the meter 5 Connect the Calibration Cable to the device or circuit under test 6 Use the arrow...

Page 7: ... Remove and replace the batteries observing polarity 3 Replace and secure the battery cover All EU users are legally bound by the Battery Ordinance to return all used batteries to community collection points or wherever batteries accumulators are sold Disposal in household trash or refuse is prohibited Disposal Follow the valid legal stipulations in respect of the disposal of the device at the end...

Page 8: ...Temperature 4 o F to 140 o F 20 o C to 60 o C Operating Humidity Max 80 up to 87ºF 31ºC decreasing linearly to 50 at 104ºF 40ºC Storage Humidity 80 Operating Altitude 7000ft 2000meters maximum Dimensions 159 x 80 x 44 mm 6 3 x 3 2 x 1 7 Weight 232g 8 2 oz without batteries Range Specifications Mode Function Range Resolution Accuracy of reading Current 0 to 50mA 0 01mA Measure Percent 25 to 230 0 1...

Reviews: