Extech Instruments PH220A User Manual Download Page 4

Model PH220 Version 3.1 11/2005 

Calibration (pH only) 

The sensitivity of a pH electrode degrades over time, it is very important to frequently 
calibrate the meter to the electrode. pH electrodes can be thought of as batteries in that 
they run down with time and use. 

 

Calibration is performed using standard buffer solutions at room temperature (78°F, 25°C). 
Buffer solutions are solutions of a known pH that have the useful property of resisting 
changes in pH. These can be purchased already mixed or in capsule form and are usually 
available in three pH values, pH 4.00, pH 7.00 and pH 10.00. 
 
Calibration is always performed using the pH 7.00 buffer first, followed by a second 
calibration. Select pH 4 buffer for the second calibration if typical measurements will be 
less than pH7 or use a pH 10 buffer if typical measurements will be greater than ph 7. 
 

In order to obtain the most accurate results possible, a pH meter should be 
calibrated at least once each day. Note that the mV (ORP) function does not require 
calibration. 

 

pH Calibration Procedure 

1. 

Place the electrode into a 7.00 pH buffer solution. Press and hold the 

CAL

 button until 

“CAL” appears in the display.  

2. 

The instrument automatically recognizes the solution and calibrates to that value. 
Note that if the solution is off by more than 1 pH unit from the 4, 7, or 10 pH buffer, or 
if the electrode slope is low, the instrument will assume an error and abort the 
calibration (End will be displayed, and the unit will return to the normal measurement 
mode.) 

3. 

During calibration, the pH reading flashes on the main display. 

4. 

When calibration is complete, the instrument automatically displays ‘SA’ followed by 
‘End’ and then returns to the normal operation mode. 

5. 

For a two or three point calibration, repeat steps 1-4. When doing a 2 or 3 point 
calibration, calibrate with pH 7 buffer first, then follow with pH 4 and/or pH 10. 

 

CAL Reminder Display 

When in pH measurement mode, the CAL icon will appear if a calibration has not been 
performed after 15 on/off cycles of the instrument. The CAL display is simply a reminder 
and will turn off when the pH electrode is recalibrated. The reminder does not affect 
function in any way. 

 

Summary of Contents for PH220A

Page 1: ...evice with tactile push buttons is battery operated and ideal for field use The PH220 simultaneously displays pH and temperature The rubber holster which protects the meter has a magnetic backing for...

Page 2: ...simultaneous pH and temperature readings Over range or open indication Displays HHH for over range or open input conditions Displays LLL for under range pH Calibration points 4 00 7 00 10 00 1 2 or 3...

Page 3: ...play icons are used and are discussed in the appropriate sections of this manual Order of Operations 1 Calibrate the instrument as described later in this manual 2 Determine the test type pH or ORP an...

Page 4: ...nction does not require calibration pH Calibration Procedure 1 Place the electrode into a 7 00 pH buffer solution Press and hold the CAL button until CAL appears in the display 2 The instrument automa...

Page 5: ...g the on off button Wait for the self test SLF tSt to finish 4 Use the MODE button to select mV mode The display will reflect the unit of measure 5 Place the electrode into a sample solution When the...

Page 6: ...time the recall mode is accessed 3 While in the RECALL mode use the STORE RECALL button to scroll through the stored readings The most recent readings will show first LAST IN FIRST OUT 4 Press and ho...

Page 7: ...it over the top of the meter ensure that the electrode is disconnected 3 Pry the battery compartment up from the top of the meter using a screwdriver or coin at one of the two indentations see diagra...

Page 8: ...ine measurement Soak in 0 1 M HCL overnight Deteriorated Gel layer Replace electrode Protein coating on electrode surface Soak in 1 gm Pepsin dissolved in 100ml of 0 1 m HCL for 30 minutes or as neede...

Page 9: ...o sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 ext 210 for authori...

Reviews: