Condensation Mode 

The Condensation feature alerts the user when the surface temperature as measured by the IR 
thermometer is close to or has reached the Dew Point temperature. 

1.  Press the power 

 button to turn the meter on. 

2.  Simultaneously press the MOIST and RH buttons. The “COND” icon will appear. 
3.  Point the meter at a surface, press the IRT button to measure the surface temperature. The 

small display will indicate the IR surface temperature and the large display will indicate the 
difference between the IR temperature and the Dew Point temperature. 

4.  The meter will then report the potential for condensation on that surface in the following 

manner 

  If the temperature of the IRT is more than 14°C (25°F) above the Dew Point, the 

temperature difference shall be displayed, with no other warning. 

  If the temperature of the IRT is 3-14°C (5-25°F) above the Dew Point, the temperature 

difference shall be displayed, along with a standard Condensation Indicator icon.  The meter 
shall beep once to confirm that the reading is in the risk area. 

  If the temperature of the IRT is less than 3°C (5°F) above the Dew Point, the temperature 

difference shall be displayed, along with a flashing Condensation Indicator icon.  The meter 
shall beep twice to confirm that the reading is in the high-risk area. 

5. 

Press the RH button to exit the mode. 

Vapor Pressure Mode 

Ambient Vapor Pressure 

1.  With the Condensation mode active, press the MODE button to display the Vapor Pressure in 

mBAR (°F) or kPa (°C). Press the 

or 

button to toggle between mBAR or kPa. 

2.  Press the MODE button to exit the Vapor Pressure mode. 

Surface Vapor Pressure 

1. 

Enter the Vapor Pressure mode as described above. 

2. 

Press the IRT button and aim the laser pointer at the surface to be measured  to display the 
Surface Vapor Pressure in mBAR (°F) or kPa (°C). 

Alarm High and Low Limit Setting 

High and Low alarm points can be set for Humidity and Moisture measurements. 
Humidity Alarm Set Procedure: 

1.  With RH% displayed, simultaneously press the RH and MODE buttons. 
2.  The “HIGH” icon will appear on the display. 
3. Press 

the 

or 

button to set the high limit desired. 

4.  Press the STORE/ALARM SET button to save the value and proceed to the LOW set value. 
5.  With the “LOW” icon in the display, Press the 

or 

button to set the low limit desired. 

6.  Press the STORE/ALARM SET button to save the value and to return to the normal mode. 
7.  If the humidity measurement is lower than the low alarm setting or higher than the high alarm 

setting, the meter will beep once every second. 

 

Shop for Extech products at:

1.877.766.5412

ShopExtech

.com

Summary of Contents for MO295

Page 1: ...User s Guide Pinless Moisture Humidity Meter with Memory IR Thermometer Model MO295 STORE MEM 01 RH MO295 GPP mBAR Shop for Extech products at 1 877 766 5412 ShopExtech com...

Page 2: ...2 Laser pointer 3 Humidity sensor 4 Temperature sensor 5 LCD display 6 Relative Humidity button 7 Mode Zero Button 8 IR thermometer button 9 Alarm set button 10 Alarm adjust down button 11 ON OFF pow...

Page 3: ...ance to spot ratio with 0 95 fixed emissivity Built in Humidity Temperature probe measures Relative Humidity Air Temperature plus Grains Per Pound GPP and Dew Point DP Ambient and Surface Vapor Pressu...

Page 4: ...rements 1 Press the power button to turn the meter on 2 Press the MOIST button to select Moisture measurement MOIST and INT internal pinless sensor will appear in the display 3 Hold the meter so that...

Page 5: ...er will display the minimum temperature measured and will update only when a lower temperature is measured 3 Press the MODE button twice to enable the MAX mode and proceed as stated above for the maxi...

Page 6: ...a flashing Condensation Indicator icon The meter shall beep twice to confirm that the reading is in the high risk area 5 Press the RH button to exit the mode Vapor Pressure Mode Ambient Vapor Pressure...

Page 7: ...l the unit beeps to save a reading into the internal memory The numeric display above the MEM icon will indicate the memory location the reading is stored to 2 When the 20 memory locations are full th...

Page 8: ...to 100 C 22 to 199 F Mixing Ratio 0 to 160g kg 0 999GPP Sample Rate 2 per second Backlight White LED Memory 20 point memory Operating Temperature 4 to 43 C 40 to 110 F Storage Temperature 30 to 60 C 1...

Reviews: