MA640-en-GB_V2.4   5/15 

Frequency and Duty Ratio Measurements 

1. 

Rotate the function switch to the 

Hz

 Position. 

2. Press 

the 

MODE

 button to select AC Voltage 

3. Press 

the 

HZ %

 button to select HZ 

4. 

Insert the black test lead banana plug into the negative 

COM

 jack. 

Insert the red test lead banana plug into the positive 

Hz

 jack. 

5. 

Touch the black test probe tip to one side of the device.  
Touch the red test probe tip to the other side of the device. 

6. 

Read the Frequency value on the display. 

7. Press 

the 

HZ %

 button to select % 

8. 

Read the Duty Ratio on the display. 

 

 

 

Type K Temperature Measurements  

1. 

Rotate the function switch to the 

Temp 

position. 

2. 

Press the MODE button to select °F or °C. 

3.  Insert the Temperature Probe into the input jacks. 
4.  Place the temperature probe tip where needed.  
5.  Read the temperature on the display.  

Note: 

In case of an open input or a temperature overrange, the meter will display 

OL”

.

 

 
 
 
 
 

Data Hold 

To freeze the LCD reading, press the 

HOLD

 button. While data hold is active, the 

HOLD

 icon 

appears on the LCD. Press the

 HOLD

 button again to return to normal operation. 

MAX/MIN  

1. Press 

the 

MAX/MIN

 button to activate the MAX/MIN recording mode. The display icon "

MAX

will appear. The meter will begins recording and displaying the maximum value measured. 

2. Press 

the 

MAX/MIN

 button and “

MIN

” will appear. The meter will display the minimum value 

measured during the recording session. 

3.  To exit MAX/MIN mode press and hold the 

MAX/MIN

 button for 2 seconds. 

 

Summary of Contents for MA640

Page 1: ...User s Guide 600Amp AC DC True RMS Clamp Meter Model MA640...

Page 2: ...injury This CAUTION symbol indicates a potentially hazardous situation which if not avoided may result damage to the product PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY OVERVOLTAGE CATEGORY I Equip...

Page 3: ...ore replacing the battery or fuses Inspect the condition of the test leads and the meter itself for any damage before operating the meter Repair or replace any damage before use Use great care when ma...

Page 4: ...RELATIVE Button 13 Battery compartment rear 14 Positive Input jack Display icons Description HOLD Data Hold AUTO Autoranging DC Direct Current AC Alternating Current MAX Max reading MIN Min reading L...

Page 5: ...tip along a length of the cord to assure placing the tip in close proximity to the live conductor NOTE The detector is designed with high sensitivity Static electricity or other sources of energy may...

Page 6: ...e side of the circuit 5 Read the voltage value in the display Resistance Measurements Note Remove power from the device under test before making resistance measurements 1 Rotate the function switch to...

Page 7: ...ay OL the device is open If both readings are very small or 0 the device is shorted Capacitance Measurements WARNING To avoid electric shock discharge the capacitor before measuring 1 Rotate the funct...

Page 8: ...position 2 Press the MODE button to select F or C 3 Insert the Temperature Probe into the input jacks 4 Place the temperature probe tip where needed 5 Read the temperature on the display Note In case...

Page 9: ...seconds LCD Backlight The LCD is equipped with backlighting for easier viewing especially in dimly lit areas Press and hold the HOLD button for 2 seconds to turn the backlight on The backlight will au...

Page 10: ...re remove the battery and store it separately Battery Replacement 1 Remove the Phillips head screws that secures the rear battery door 2 Open the battery compartment 3 Replace the 9V battery 4 Secure...

Page 11: ...0 0 VDC 0 1V Resistance 600 0 0 1 1 0 4 digits 6 000k 0 001k 1 5 2 digits 60 00k 0 01k 600 0k 0 1k 6 000M 0 001M 2 5 3 digits 60 00M 0 01M 3 5 5 digits Capacitance 40 00nF 0 01nF 4 0 20 digits 400 0nF...

Page 12: ...ranges 50 60Hz and 5 to 100 of range Operating Temperature 5 C to 40 C 41 F to 104 F Storage Temperature 20 C to 60 C 4 F to 140 F Operating Humidity Max 80 up to 31 C 87 F decreasing linearly to 50...

Reviews: