MA160
‐
en
‐
GB_V1.0
10/15
10
Voltage
Measurements
WARNING:
Remove
the
test
lead
probe
covers
for
CAT
II
1000V
installations.
Use
the
test
lead
probe
covers
for
CAT
III
600V
installations.
Do
not
measure
voltages
greater
than
600VAC
or
1000VDC.
CAUTION
:
When
connecting
the
test
leads
to
the
circuit
or
device
under
test,
connect
the
black
lead
before
the
red;
when
removing
the
test
leads,
remove
the
red
before
the
black
lead.
Disconnect
probes
and
circuit
under
test
after
all
measurements
are
completed.
AC
Voltage
Measurements
1.
Insert
the
black
test
lead
banana
plug
into
the
negative
(COM)
jack
and
the
red
test
lead
banana
plug
into
the
positive
(V/
Ω
)
jack.
2.
Turn
the
function
switch
to
the
position.
Use
the
M
button
to
select
AC.
3.
Read
the
Warning
and
Caution
statements
above
to
determine
whether
or
not
to
use
the
test
lead
probe
covers.
4.
Touch
the
test
probe
tips
to
the
circuit
under
test.
5.
Read
the
digital
value
shown
on
the
display.
The
meter
is
Auto
Ranging
and
therefore
selects
the
proper
decimal
point
position.
The
meter
also
displays
the
measurement
type,
unit
symbols,
and
other
relevant
multifunction
icons.
6.
The
meter
is
capable
of
detecting
AC
voltages
to
600V.
Fig
4
‐
1
AC
VOLTAGE
MEASUREMENTS
Shop for Extech products at:
1.888.610.7664
ShopExtech
.com