Operation 

Setup 

1.

Press the

 Power button to turn the meter on. 

2.

The meter will turn on in the default mode: AUTO LCR and 1kHz

3.

APO will be active with auto shutoff every 5 minutes unless an AC adaptor is used.

LCR primary parameter selection 

1.

The meter will turn on with auto parameter selection enabled and with the ”APO” (auto power
off), “Auto” (autoranging) and “LCR” (auto-parameter) icons appearing at the top of the 
display. 

2.

To manually select the parameter, press the L/C/R button to step through and select the
parameter needed. Each button press will sequentially display: 

Auto-LCR 

Auto Range 

Auto Parameter 

Auto-L Auto 

Range  Inductance 

Auto-C Auto 

Range  Capacitance 

Auto-R Auto 

Range  Resistance 

DCR

DC 

Resistance

3.

The value of the primary parameter will be displayed in the upper display and the secondary
parameter in the lower display. 

D/Q/

θ

 secondary parameter selection 

1.

The meter will turn on with auto parameter selection “LCR” enabled. .The primary and
secondary parameter will be automatically selected based on the value of the measured 
impedance. 

2.

To manually select the secondary display, first select the primary display.

3.

Press the D/Q/

θ

 button to select the secondary parameter:

D, Q, ESR(RP) or 

θ

 

D, Q, ESR(RP) or 

θ

 

R

none

DCR

none

4.

The value of the secondary parameter will be displayed in the lower display

Series or Parallel 

1.

When the L/C/R function mode is selected and ‘AUTO” is enabled, the default measurement
in serial or parallel mode is auto selected. A parallel equivalent circuit (Lp, Cp or Rp) will be 
selected if the impedance is greater than 10k

. A series equivalent circuit ( Ls, Cs or Rs) will 

be selected if the impedance is less than 10k

.  

2.

Press the SER/PAR button as needed to change the default selection.

Frequency 

Press the FREQ button to change the test frequency. The selections are: 100Hz, 120Hz, 1kHz, 
10kHz and 100kHz. 

Data Hold 

Press the HOLD button to freeze the reading in the display. Push the HOLD key again to cancel 
the hold mode and return to normal operation. 

actoolsupply.com

actoolsupply.com

Summary of Contents for lcr200

Page 1: ...User s Guide Digital LCR Meter Model LCR200 D actoolsupply com actoolsupply com actoolsupply com Extech LCR200 Passive Component LCR Meter...

Page 2: ...l equivalent circuit This meter is shipped fully tested and calibrated and with proper use will provide years of reliable service Features 19 999 1 999 count dual LCD display Auto LCR smart check and...

Page 3: ...Button 9 Input terminals pin terminals 10 Input terminals banana terminals 11 Hold Button 12 RS232 Port not active on this model 13 L C R Button 14 Enter Button 15 SER PAR Button 16 Sort Button 17 Se...

Page 4: ...n with auto parameter selection LCR enabled The primary and secondary parameter will be automatically selected based on the value of the measured impedance 2 To manually select the secondary display f...

Page 5: ...in the display press the SORT button to establish the reference value The main display will indicate PASS and the lower display will indicate the value of the component The default sorting is 1 Note I...

Page 6: ...de 7 If PASS appeared for both the OPEN and SHORT modes the calibration data will be stored 8 If FAIL appeared for either calibration the impedance was too large to zero out and the data is not stored...

Page 7: ...e two screws holding the rear bottom battery cover 2 Remove and replace the 9V battery 3 Secure the cover with the two screws All EU users are legally bound by the battery ordinance to return all used...

Page 8: ...88 x 41mm 7 6 x 3 5 x 1 6 Weight 385 g 13 6 oz meter only Electrical Specifications 23 5 C DC Resistance Range Accuracy Remark 20 0 5 rdg 5digits After calibration 200 0 5 rdg 5digits 2k 0 5 rdg 5dig...

Page 9: ...calibration 20nF 0 5 rdg 5digits 0 5 rdg 5digits 200nF 0 5 rdg 5digits 0 5 rdg 5digits 2000nF 0 5 rdg 5digits 0 5 rdg 5digits 20uF 0 5 rdg 5digits 0 5 rdg 5digits 200uF 0 5 rdg 5digits 0 5 rdg 5digit...

Page 10: ...5d 0 5 5d 20H 0 5 5d 0 5 5d 200H 0 5 5d 0 8 5d After calibration 2000H 1 5d After calibration Range Accuracy Accuracy Remark 10kHz 100khz 20uH 1 5d 1 5d After calibration 200uH 0 8 5d 0 8 5d After cal...

Reviews: