Inductance ( Lp/Ls ) : D
≦
0.1
Range
Accuracy
Accuracy
Remark
100Hz/120Hz
1000Hz
20uH
± (1% + 5d )
± (1% + 5d )
After calibration
200uH
± (1% + 5d )
± (1% + 5d )
After calibration
2000uH
± (0.8% + 5d )
± (0.8% + 5d )
20mH
± (0.5% + 5d )
± (0.5% + 5d )
200mH
± (0.5% + 5d )
± (0.5% + 5d )
2000mH
± (0.5% + 5d )
± (0.5% + 5d )
20H
± (0.5% + 5d )
± (0.5% + 5d )
200H
± (0.5% + 5d )
± (0.8% + 5d )
After calibration
2000H
± (1% + 5d )
-----------------------
After calibration
Range
Accuracy
Accuracy
Remark
10kHz
100khz
20uH
± (1% + 5d )
± (1% + 5d )
After calibration
200uH
± (0.8% + 5d )
± (0.8% + 5d )
After calibration
2000uH
± (0.5% + 5d )
± (0.5% + 5d )
20mH
± (0.5% + 5d )
± (0.5% + 5d )
200mH
± (0.5% + 5d )
-----------------------
2000mH
± (0.5% + 5d )
-----------------------
Note:
* If the impedance is larger than 10k
Ω
, Lp is shown on the display.
* If the impedance is less than 10k
Ω
, Ls is shown on the display.