background image

6

EC500-en-US_v2.9 7/20

Recalling Stored Readings 

1.  Press the 

CAL/RECALL

 button and then press the 

MODE/HOLD

 button. A location number (1 

through 25) will briefly appear and then the value stored in that location will appear. The 
displayed units will flash, indicating that the storage recall mode is active. 

2.  The last stored reading will be displayed first. Pressing and releasing the 

MODE/HOLD

 button 

will scroll through the stored readings one at a time. The location number is displayed first, 
followed by the reading stored in that location. 

3.  To exit the storage mode, press the 

CAL/RECALL

 button and the meter will return to normal 

operation

nd . 

Clear Stored Memory 

clr

memory is cleared. 

Changing Temperature Units 

To change the displayed temperature units (

o

C or 

o

F): 

1.  With the unit OFF, press and hold down the 

CAL/RECALL

 button. 

2.  With the 

CAL/RECALL

 button depressed momentarily press the 

ON/OFF

 button. 

CAL/RECALL

 button. The unit will power on with 

temperature displayed in the new units. 

Data Hold Mode 

Press the 

MODE/HOLD

 button to hold (freeze) a reading in the display. The meter will enter the 

 

Note: This also stores the reading.  
Press the 

MODE/HOLD

 button again to return to normal operation. 

Auto Power OFF 

The auto power off feature automatically shuts the meter off 10 minutes after the most recent button 
press. 

Auto Power OFF Disable 

To disable the Auto Power Off feature:  
1. 

Turn the unit on 

2. 

Press 

CAL/RECALL

 once (

Quickly

3. 

Immediately and simultaneously press the 

MODE/HOLD

 and 

ON/OFF

 buttons for 

approximately 2 seconds, until 

oFF

 is briefly displayed 

To disengage this feature, turn the unit off with the 

ON/OFF

 button. The next time the unit is 

powered up, Auto Power OFF mode will be engaged again. 

Low Battery Indication  

on will appear in the display. Refer to the 

Maintenance section for battery replacement information. 

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for ExStik EC500

Page 1: ...User Manual ExStik EC500 pH Conductivity TDS Salinity Temperature Meter www com information itm com 1 800 561 8187 ...

Page 2: ...on and conductivity electrodes Before the first use or after storage soak the electrode in tap water or pH 4 buffer solution for about 10 minutes White KCL crystals may be present in the cap or on the electrode This is to be expected depending on the length of time in storage These crystals will dissolve while soaking the electrode or they can be rinsed away with tap water For best results calibra...

Page 3: ... RECALL button 5 ON OFF button 6 Electrode Collar 7 pH Conductivity Electrode Note The Electrode cap is not shown LCD Display 1 Bargraph display 2 Measurement units 3 Main display 4 Range calibration and low battery indicators 5 Temperature display 6 Renew indicator 7 Reading hold indicator www com information itm com 1 800 561 8187 ...

Page 4: ...e Measurement 1 Press the ON button display during the turn on diagnostics 2 Depress and hold the MODE HOLD key to scroll to the desired measurement mode 3 Insert the electrode into the sample making sure that the electrodes are completely submersed 4 Slowly stir the solution with the electrode to remove air bubbles if in the Conductivity TDS or Salinity mode 5 If in the Conductivity TDS or Salini...

Page 5: ...rned on or when changing measurement function to TDS Note In the Salinity mode the ratio is 0 4 to 0 6 auto To change the ratio while in the TDS measurement mode ppm or mg l 1 Press and release the CAL RECALL button twice in succession The stored ratio will appear in the display 2 Press the MODE HOLD button to increase the ratio value in steps of 0 1 3 When the desired ratio is displayed press and...

Page 6: ... the CAL RECALL button depressed momentarily press the ON OFF button CAL RECALL button The unit will power on with temperature displayed in the new units Data Hold Mode Press the MODE HOLD button to hold freeze a reading in the display The meter will enter the Note This also stores the reading Press the MODE HOLD button again to return to normal operation Auto Power OFF The auto power off feature ...

Page 7: ...minder to calibrate pH and will turn off when the pH electrode is recalibrated The reminder does not affect function in any way RENEW Display is not performing to expected specifications If cleaning and recalibration does not cause the RENEW icon to disappear replace the probe see optional accessories on the last page of this manual The RENEW display appears as a result of the pH electrode slope f...

Page 8: ...nges keep the existing calibration data just the symbols are removed Calibration of all three ranges must be performed during one power on period for all three range calibration symbols to appear See Reset Calibration Data to clear all calibration data from the meter Note The meter allows for a 1 2 or 3 point calibration If calibration is done for more than one point the lowest value standard shou...

Page 9: ...de Press and hold the ON OFF key for 4 seconds Change Measurement Mode On Any Press and hold the MODE HOLD key for at least 2 seconds the modes will scroll by until the key is released Enter Cond TDS Ratio On TDS ppm or mg l Press and release the CAL RECALL key twice in quick succession Change Cond TDS Ratio On TDS ratio Momentary press of the MODE HOLD key each key press increases the ratio by 0 ...

Page 10: ...Ratio 0 4 to 0 6 auto Conductivity ATC 2 0 per o C Conductivity ATC Range 32 0 F to 140o F 0 0 C to 60 0o C Temperature Range 23 0 F to 194o F 5 0 C to 90 0o C Temperature Resolution 0 1 up to 99 9 1 100 Temperature Accuracy 1 8 F 1 C from 23 to 122 F 5 to 50 C 5 4 F 3 C from122 to 194 F 50 to 90 C Accuracy Conductivity 2 full scale TDS 2 full scale Salinity 2 full scale Measurement Memory 25 tagg...

Page 11: ...is not to be used for 60 days or more remove the battery and store separately Electrode Replacement 1 To remove an electrode unscrew and completely remove the electrode collar turn the collar counter clockwise to remove 2 Gently rock the electrode from side to side pulling it downwards until it disconnects from the meter 3 To attach an electrode carefully plug the electrode into the meter socket n...

Page 12: ... 4 or 7 buffer for 1 hour Heavy grease Oil Alcohol Maximum of 5 minute soak scrub with a soft brush Rinse thoroughly with DI water recondition in 4 or 7 buffer for 1 hour Lime and hydroxide coatings 10 acetic acid Soak until coating dissolved maximum of 5 minutes Rinse thoroughly with DI water recondition in 4 or 7 buffer for 1 hour Please Note Since the EC500 does not have a refillable pH referen...

Page 13: ...ards Unit will not calibrate in conductivity mode Dirty probe Clean conductivity probe See cleaning instructions Unit will not calibrate in conductivity mode Damaged conductivity probe Replace electrode see reorder information Unit will not calibrate in conductivity mode Trapped air bubbles Tap or stir to release air bubbles Unit will not turn on Batteries are low or dead Replace batteries Unit wi...

Page 14: ...rand products we sell We offer NIST traceable calibration for most of our products Contact us for information on calibration and repair availability refer to the contact information below Annual calibrations should be performed to verify meter performance and accuracy Product specifications are subject to change without notice Please visit our website for the most up to date product information Co...

Reviews: