12
EC500-en-US_v2.9 7/20
Cleaning Recommendations
When cleaning the probe, take care not to scratch or damage the sensing surface or the platinized
electrode surfaces.
Contaminant
Cleaning Solution
Instructions
Water soluble substances
Deionized water
Soak or scrub with a soft
brush. Recondition in 4 or 7
buffer for 1 hour.
Grease & Oil
Warm water and household
detergent
Soak or scrub with a soft
brush, maximum of 10
minutes. Rinse thoroughly
with DI water, recondition in 4
or 7 buffer for 1 hour.
Heavy grease & Oil
Alcohol
Maximum of 5 minute soak,
scrub with a soft brush. Rinse
thoroughly with DI water,
recondition in 4 or 7 buffer for
1 hour.
Lime and hydroxide coatings
10% acetic acid
Soak until coating dissolved,
maximum of 5 minutes. Rinse
thoroughly with DI water,
recondition in 4 or 7 buffer for
1 hour.
Please Note:
Since the EC500 does not have a refillable pH reference electrolyte chamber, it is
important not to soak the electrode in the above solutions for more than the recommended times. To
do so may cause a reference potential shift, which will cause degradation in performance or failure.