CL500-en-GB_V1.2   8/15 

3

Getting

 

Started

 

Equipment

 

Required

 

A

 

chlorine

 

test

 

requires

 

a

 

10ml

 

sample

 

bottle,

 

reagent

 

powder

 

for

 

testing

 

Free

 

and

 

Total

 

chlorine,

 

and

 

the

 

CL500

 

meter.

 

Measurement

 

Procedure

 

Calibration

 

considerations

 

1.

 

Ensure

 

that

 

the

 

meter

 

has

 

been

 

calibrated

 

prior

 

to

 

use.

 

2.

 

Refer

 

to

 

the

 

calibration

 

section

 

for

 

this

 

procedure.

 

 

Test

 

sample

 

“Zero”

 

measurement

 

Note:

 

Prior

 

to

 

testing,

 

set

 

the

 

meter

 

mode

 

to

 

FREE

 

or

 

TOTAL

 

based

 

on

 

the

 

desired

 

test

 

method.

  

See

 

Mode

 

Selection.

 

1.

 

The

 

meter

 

should

 

be

 

in

 

the

 

OFF

 

position.

 

2.

 

Place

 

the

 

test

 

sample

 

into

 

the

 

sample

 

bottle

 

up

 

to

 

the

 

10mL

 

line

 

(minimum).

 

Place

 

the

 

cap

 

on

 

the

 

bottle

 

and

 

tighten.

 

3.

 

Clean

 

and

 

dry

 

the

 

outside

 

of

 

the

 

sample

 

bottle

 

to

 

ensure

 

a

 

clean

 

surface.

 

4.

 

Open

 

the

 

cover

 

of

 

the

 

test

 

chamber

 

and

 

insert

 

the

 

bottle

 

completely

 

into

 

the

 

chamber.

 

5.

 

Line

 

up

 

the

 

white

 

mark

 

on

 

the

 

bottle

 

with

 

the

 

white

 

mark

 

of

 

the

 

test

 

chamber.

 

6.

 

Close

 

the

 

cover

 

and

 

lock

 

it

 

in

 

place.

 

7.

 

Press

 

the

 

“POWER”

 

button

 

to

 

turn

 

the

 

meter

 

ON.

  

The

 

display

 

will

 

show

 

‘Free’

 

or

 

‘Total’

 

for

 

1

 

second

 

and

 

then

 

‘CAL0’.

 

The

 

meter

 

is

 

now

 

ready

 

for

 

the

 

“zero”

 

test.

 

8.

 

Press

 

the

 

“Zero”

 

button,

 

the

 

display

 

will

 

flash

 

“tEST”

 

and

 

then

 

“0.00”ppm.

  

The

 

Zero

 

Test

 

is

 

now

 

complete.

 

Do

 

not

 

power

 

down

 

the

 

meter.

 

Measurement

 

of

 

Test

 

Solution

 

1.

 

Open

 

the

 

test

 

chamber

 

and

 

remove

 

the

 

test

 

bottle.

 

2.

 

Add

 

the

 

required

 

DPD

 

powder

 

(Free

 

or

 

Total)

 

into

 

the

 

sample

 

and

 

then

 

shake

 

for

 

20

 

seconds

 

until

 

combined.

 

Place

 

the

 

cap

 

on

 

the

 

bottle

 

and

 

tighten.

 

For

 

the

 

Free

 

Chlorine

 

test,

 

use

 

the

 

Free

 

Chlorine

 

DPD

 

powder

 

For

 

the

 

Total

 

Chlorine

 

test,

 

use

 

the

 

Total

 

Chlorine

 

DPD

 

powder

 

3.

 

Clean

 

and

 

dry

 

the

 

outside

 

of

 

the

 

sample

 

bottle

 

to

 

ensure

 

a

 

clean

 

surface.

 

4.

 

Wait

 

2

 

minute

 

before

 

placing

 

the

 

test

 

bottle

 

into

 

the

 

meter

 

to

 

allow

 

time

 

for

 

the

 

solution

 

to

 

mix

 

thoroughly.

 

5.

 

Open

 

the

 

cover

 

of

 

the

 

test

 

chamber

 

and

 

insert

 

the

 

bottle

 

completely

 

into

 

the

 

chamber.

 

6.

 

Line

 

up

 

the

 

white

 

mark

 

on

 

the

 

bottle

 

with

 

the

 

white

 

mark

 

of

 

the

 

test

 

chamber.

 

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for CL500

Page 1: ...USER GUIDE Free and Total Chlorine Tester Patent Pending Model CL500 www com information itm com 1 800 561 8187 ...

Page 2: ...ease visit our website www extech com to check for the latest version of this User Guide Product Updates and Customer Support Applications The CL500 can be used wherever a measurement of the total chlorine in water as needed This is also referred to as total residual chlorine It measures the total of all chlorine present in all forms including dissolved free chlorine chloramines hypochlorous acid ...

Page 3: ...chamber 6 Close the cover and lock it in place 7 Press the POWER button to turn the meter ON The display will show Free or Total for 1 second and then CAL0 The meter is now ready for the zero test 8 Press the Zero button the display will flash tEST and then 0 00 ppm The Zero Test is now complete Do not power down the meter Measurement of Test Solution 1 Open the test chamber and remove the test bo...

Page 4: ...ode momentarily press the Rec button The display will show REC MAX Momentarily press the HOLD button the display will now show REC The Max value has been erased 6 To erase the Min value and continue recording while in the record mode momentarily press the Rec button The display will show REC MAX Press the REC button once again and now the display will show REC MIN Momentarily press the HOLD button...

Page 5: ...ease the CAL button 4 Clean the sides of the Zero Chlorine solution bottle and insert it into the testing chamber 5 Line up the white mark on the bottle with the white mark in the chamber and close the cover and latch it 6 Press CAL once more and the display will flash tEST and show the following 7 The meter is now ready for zero chlorine calibration 8 Press CAL button once The LCD will flash CAL ...

Page 6: ...rk on the bottle with the white mark in the chamber close and latch the cover 3 Press CAL once more the display will flash CAL and then show the following 4 When calibration is complete the LCD will return to the normal operating mode and measure the test solution The display should show 1 00 ppm 5 The meter is now ready to measure test solutions CAL 1 00 ppm www com information itm com 1 800 561 ...

Page 7: ...rechargeable batteries in household waste As consumers users are legally required to take used batteries to appropriate collection sites the retail store where the batteries were purchased or wherever batteries are sold Disposal Do not dispose of this instrument in household waste The user is obligated to take end of life devices to a designated collection point for the disposal of electrical and ...

Page 8: ...m Sample Time Approximately 1 second Power Six 1 5 V batteries Power Consumption Stand By Mode approx 4mADC Testing Mode approx 12mADC Dimensions 155 x 76 x 62mm 6 1 x 3 0 x 2 4 320g 0 70lbs Auto Power OFF Automatically switches OFF 10 minutes after the last key press Method The measuring method is an adaptation of the USEPA method 330 5 and standard method 4500 Cl G Copyright 2014 2015 FLIR Syste...