Programming Mode P2; Datalogger Start Modes
1. Press the SET button to access the setup mode
2. Use the arrow buttons to select parameter P2
3. Press the SET button again
4. Select one of the four Start modes using the up/down arrow buttons (see list and diagram
below)
a.
rEp
(Repeat Start mode): Logger starts
once readings are downloaded to PC
b.
hEy
(Keypad Start): Datalogging begins
after the START/STOP button is held for
2 seconds
c.
imm
(Immediate Start): Recording starts
immediately after exiting the setup mode
d.
SCh
(Scheduled Start): Logging begins
at a preset date and time
5. Press the SET button to save the setting