42280-EU-EN-V4.4-12/11
5
Manual Programming
The Model 42280 has several programming modes (P1, P2, P3, P4, P5, and P6). These
parameters must be programmed before the Datalogger or the Alarm can be used.
List of Programming Modes
P1
- Number of samples to record (1, 2, 4, 8, 12, or 16 x 1000; 1=1000, 2=2000, etc.)
P2
- Datalogging Start Methods (Immediate, Keypad, Scheduled, and Repeat modes)
P3
– Start Time (set the time and date when Datalogging will automatically begin); Note that P3
will not appear in the menu if ‘SCh’ is not selected in P2.
P4
- Sample Rate (set the datalogging interval from 1 second up to 12 hours)
P5
- Alarm Mode (set High and Low Temperature/Humidity limits)
P6
- Real Time Clock (set the meter’s internal date and time clock)
Programming Mode P1; Number of readings (sample points) to record
1.
Press the SET button to access the setup mode
2.
Use the arrow buttons to select parameter P1 (see sample display below)
3.
Press the SET button again
4.
Use the up/down arrow buttons to select the
desired number of samples (in thousands) to record
for a datalogging session (1, 2, 4, 8, 12, or 16;
where 1=1000, 2=2000, 4=4000, and so on)
5.
Press the SET button to save setting
6.
Use the arrow buttons to select another programming mode or press MODE to exit the
programming mode.