Model 421504 Version 1.9 December 2004 

Temperature units – Degrees 

o

C or 

o

F

 

The Primary and Second display (one for each RTD input) can read in either degrees 

o

C or 

o

F. Press the C/F button to toggle between 

o

C and 

o

F temperature units. The meter 

remembers the selected unit of measure the next time the meter is used. 

Data Hold 

To freeze the LCD display, press the HOLD button. The HOLD icon will appear on the 
upper right-hand side of the LCD. Press the HOLD button again to return to normal 
operation (the HOLD display icon will switch off). 

MIN, MAX, AVG Function with Timer (for Primary display only) 

In the MIN/MAX/AVG mode of operation the meter tracks the highest (MAX), lowest (MIN), 
and average (AVG) primary display readings for easy recall. 

1.  To enter this mode, press the MIN/MAX button (the RECORD display icon will appear). 

Use the MIN/MAX button to cycle through MIN, MAX, and AVG modes each with its 
own LCD display icon. 

2.  Each time a new MAX or MIN is detected, the previous MAX or MIN displayed value is 

replaced and an audible tone will sound. 

3.  Use the HOLD button to start and stop recording.  When recording is stopped, the 

stored values are held in meter memory until the MIN/MAX/AVG mode is exited or 
resumed (resume by pressing the HOLD button again). 

4.  The AVG mode keeps a true running average of temperatures recorded since the 

MIN/MAX button was first pressed (up to a maximum of 22 hours). If 22 hours elapse 
the average stored at this point is held on the display and is no longer updated (the 
MIN and MAX functions continue without interruption indefinitely). If an overload is 
recorded, the AVG function is halted and the AVG display shows "-------".  

5.  In this mode, the Primary displayed value is the MIN, MAX, or AVG reading depending 

upon which icon appears on the LCD. 

6.  Each value for MIN/MAX/AVG has an associated time value (bottom right) indicating 

when the displayed reading was taken. Set the timer clock in the SET mode (refer to 
the SET MODE section). 

7.  The MIN/MAX/AVG mode cancels the Auto Power OFF feature, disables most of the 

meter keys, and will not allow a manual power OFF. Press and hold the MIN/MAX key 
for several seconds to exit MIN/MAX/AVG mode and return to normal operation. 

Relative Mode (for Primary display only) 

The Relative Mode permits the user to store a temperature and compare it to subsequent 
temperature readings. In Relative mode the primary display indicates the actual 
temperature minus the relative temperature. The user can store the current primary 
temperature reading as the relative temperature or a temperature can be entered 
manually. To manually enter a relative temperature, refer to the SET MODE section.  
 
To enter the relative mode, press the REL button; The REL icon will appear on the LCD 
and the temperature display will show zero (indicating that the actual and relative 
temperatures are the same).  Press the REL key again to exit the REL Mode. 

Summary of Contents for 421504

Page 1: ...put RTD Thermometer Model 421504 Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com 99 Washington Street Melrose MA 02176 Fax 781 665 0780 TestEquipmentDepot com ...

Page 2: ...5 RTD Probe Pt 3926 100Ω 328 to 1166 o F 200 to 630 o C U S curve Alpha 003926 Temperature units Selectable C or F temperature units Accuracy is specified for operating temperatures in the range of 64 to 82 o F 18 to 28 o C for a period of one year not including the RTD probe error 0 05 rdg 0 2 o C for o C Scale Accuracy 0 05 rdg 0 4 o F for o F Scale Temperature Coefficient Multiply the applicabl...

Page 3: ...robe inputs 2 LCD multifunction display 3 Keypad 4 Protective holster Note that the tilt stand and battery compartment are located on the back of the meter Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com ...

Page 4: ...Model 421504 Version 1 9 December 2004 5 Display Description Keypad Description ...

Page 5: ... type alpha 3850 or 3926 that is connected to the T2 terminal using the Pt385 3926 button the button with the number 0 overlay 3 The LCD display indicates which alpha curve is selected Note Alpha curve 385 is the DIN IEC worldwide current standard while Alpha 3926 is an older standard If unsure check with the Probe manufacturer If still in doubt use the 385 selection The meter remembers the select...

Page 6: ... MAX button was first pressed up to a maximum of 22 hours If 22 hours elapse the average stored at this point is held on the display and is no longer updated the MIN and MAX functions continue without interruption indefinitely If an overload is recorded the AVG function is halted and the AVG display shows 5 In this mode the Primary displayed value is the MIN MAX or AVG reading depending upon which...

Page 7: ...ered as 0080 100 as 0100 etc 3 Press the ENTER button once 4 Program the LO Alarm Limit value temperature at which the meter audibly alerts the user 5 Press the ENTER button once to return to normal operation The alarms will be armed giving an audible alert when a limit is reached 6 When the actual temperature meets or exceeds the user programmed temperature limits a continuous alert tone will be ...