background image

Model 421504 Version 1.9 December 2004 

Specifications 

General Specifications 

Display 

 

5-digit LCD with overload/open-input indication

 

Over-range and open input  Dashes are displayed in place of a temperature reading 

Power Supply

 

9V battery

 

Battery life

 

100 hours typical with carbon zinc battery

 

Auto Power Off 

Meter turns off automatically after 30 minutes 

Safety 

CE approved and water resistant meter housing 

Weight      

 

11.73oz. (365 g)

 

Dimensions

 

7.56 x 3.58 x 2" (192 x 91 x 52.5mm)

 

Accessories supplied

 

9V battery, RTD connectors (2), and protective holster

 

Range Specifications 

Display Resolution

 

0.2

o

F or  0.1

o

C Resolution

 

RTD Probe (Pt 385 100

)

 

-328 to 1472

o

F (-200 to 800

o

C) 

(European curve Alpha = .00385)

 

RTD Probe (Pt 3926 100

)

 

-328 to 1166

o

F (-200 to 630

o

C) 

(U.S. curve Alpha = .003926)

 

Temperature units

 

Selectable C or F temperature units 

 

Accuracy is specified for operating temperatures in the 
range of  64 to 82

o

F (18 to 28

o

C) for a period of one 

year, not including the RTD probe error

 

±(0.05% rdg + 0.2

o

C) for 

o

C Scale

 

Accuracy

 

±(0.05% rdg + 0.4

o

F) for 

o

F Scale

 

Temperature Coefficient

 

Multiply the applicable accuracy specification per 

o

from 0 to 18

o

C and 28 to 50

o

C (32 to 64

o

F and 82 to 

122

o

F) by 0.1

 

Input Protection

 

24VDC or 24VAC rms max. input on any combination 
of inputs. 

 

NOTE: Max. Voltage between T1 and T2 during 
measurement  = 1V

 

Reading rate

 

One reading per second

 

Input connectors

 

Accepts standard 3-terminal RTD sub-miniature 
connectors (flat blade, 7.9mm spacing)

 

Ambient Operating Range

 

32 to 122

o

F (0 to 50

o

C); less than 80% RH

 

Storage Temperature

 

-4 to 140

o

F (-20 to 60

o

C); less than 70% RH

 

 
 
 

Summary of Contents for 421504

Page 1: ...put RTD Thermometer Model 421504 Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com 99 Washington Street Melrose MA 02176 Fax 781 665 0780 TestEquipmentDepot com ...

Page 2: ...5 RTD Probe Pt 3926 100Ω 328 to 1166 o F 200 to 630 o C U S curve Alpha 003926 Temperature units Selectable C or F temperature units Accuracy is specified for operating temperatures in the range of 64 to 82 o F 18 to 28 o C for a period of one year not including the RTD probe error 0 05 rdg 0 2 o C for o C Scale Accuracy 0 05 rdg 0 4 o F for o F Scale Temperature Coefficient Multiply the applicabl...

Page 3: ...robe inputs 2 LCD multifunction display 3 Keypad 4 Protective holster Note that the tilt stand and battery compartment are located on the back of the meter Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com ...

Page 4: ...Model 421504 Version 1 9 December 2004 5 Display Description Keypad Description ...

Page 5: ... type alpha 3850 or 3926 that is connected to the T2 terminal using the Pt385 3926 button the button with the number 0 overlay 3 The LCD display indicates which alpha curve is selected Note Alpha curve 385 is the DIN IEC worldwide current standard while Alpha 3926 is an older standard If unsure check with the Probe manufacturer If still in doubt use the 385 selection The meter remembers the select...

Page 6: ... MAX button was first pressed up to a maximum of 22 hours If 22 hours elapse the average stored at this point is held on the display and is no longer updated the MIN and MAX functions continue without interruption indefinitely If an overload is recorded the AVG function is halted and the AVG display shows 5 In this mode the Primary displayed value is the MIN MAX or AVG reading depending upon which...

Page 7: ...ered as 0080 100 as 0100 etc 3 Press the ENTER button once 4 Program the LO Alarm Limit value temperature at which the meter audibly alerts the user 5 Press the ENTER button once to return to normal operation The alarms will be armed giving an audible alert when a limit is reached 6 When the actual temperature meets or exceeds the user programmed temperature limits a continuous alert tone will be ...