Sound Pressure Measurements (SPL) 

In SPL mode the display shows readings of the sound pressure level.  

 
1. 

Turn the meter on by pressing the 

green on/off

 button. 

2. 

Select the SPL mode by pressing the 

Leq SEL SPL

 button until SPL appears on the LCD. 

3. 

Select the desired Frequency Weighting as needed (A or C). 
Change the Frequency Weighting by pressing the 

A/C

 button. The 'A' or 'C' icon will display on 

the right-hand area of the LCD. 

Note:

 With ‘A’ weighting selected, the meter responds like the human ear. ‘A’ weighting is 

commonly used for environmental measurements, OSHA regulatory testing, law enforcement, 
and workplace design. Select ‘C’ weighting for flat response measurements. ‘C’ weighting is 
typically used for the sound level analysis of machines, engines, etc. Most hearing 
conservation related testing is performed using 'A' Weighting. 

 
4. 

Select the desired Response Time as needed (Fast, Slow or Impulse). 
Change the Response Time by pressing the 'FAST/SLOW/IMPULSE' button. The 'FAST'  'SLOW' 
or IMP icon will display. 

Note:

 Select FAST to capture noises that occur very quickly. In FAST mode, the meter 

responds in 125ms. Select the SLOW Mode (meter responds in 1s) to monitor a sound 
source that has a reasonably consistent noise level or to average quickly changing levels.

 

In 

IMPULSE mode the meter responds in 35ms and is used to capture the rms value of 
repetitive impulse noises

.

 Selection of Fast or Slow is determined by the application and any 

directives or standards related to that application. Most hearing conservation related testing is 
performed using a SLOW Response Time setting. 

 
5. 

Hold the instrument comfortably in hand and away from your body or position it on a tripod. Point 
the microphone toward the noise source, the sound pressure level will be displayed on the 
meter’s LCD display. 

 

MIN/MAX Function 
 

Press the 

MAX MIN INST

 button to begin capturing the, Maximum (MAX) OR Minimum (MIN) sound 

values. The display will indicate the MAX

 

or MIN value (depending on which is selected) captured 

since the MAX/MIN button was pressed. Press and hold the  

MAX MIN INST

 button to return to the instantaneous reading mode 

 

Note:

  

When the meter is measuring the following buttons cannot be used: 

 

 

FAST, SLOW, IMPULSE 

 A/C 

Weighting 

 INTEG 

TIME 

 RECORD 

ERASE 

 DATE/TIME 

 

UP/DOWN Arrow keys 

 

 

Summary of Contents for 407780

Page 1: ...User s Guide Integrating Sound Level Datalogger Model 407780...

Page 2: ...MAX L and MIN L measurements Response time FAST SLOW and IMPULSE and frequency weighting A and C are programmable The analog output AC and DC can be used with headphones chart recorders or other devic...

Page 3: ...bol flashes once per second while recording Erase data Press hold the RECORD ERASE key for 3 seconds to erase data the entire LCD will flash three times 11 Real time clock with calendar 12 Measuring t...

Page 4: ...of sound pressure level 10 Data recording 11 Measuring value 12 Memory full 13 Unit 14 Frequency weighting A C 15 IMPULSE time weighting 16 When the TIME display icon switches on the real time clock...

Page 5: ...orm a manual recording select the 24 hour interval Change the run time by pressing the INTEG TIME button The run time will appear in the middle of the LCD Press the or button to select the preset time...

Page 6: ...e To perform a manual recording select the 24 hour interval Change the run time by pressing the INTEG TIME button The run time will appear in the middle of the LCD Press the or button to select the pr...

Page 7: ...to capture noises that occur very quickly In FAST mode the meter responds in 125ms Select the SLOW Mode meter responds in 1s to monitor a sound source that has a reasonably consistent noise level or t...

Page 8: ...the arrow keys to set the correct month 4 Repeat step 3 to set the day hour minute and second 5 To confirm the new date and time press the DATE TIME button Note that the 407780 has a back up battery N...

Page 9: ...th a dry cloth Do not use abrasives or solvents Analog output The 407780 has two analog output jacks located on its right side one for AC and one for DC For DC the meter transmits 10mv dB For AC the f...

Page 10: ...and the measurement time in the supplied software The meter can now record remotely disconnected from the PC at the rate selected 4 Press the RECORD ERASE key to put the meter in Record Standby The RE...

Page 11: ...ay 4dB per step over 100dB range with 50 segments Bargraph update rate 16 times per second Display warnings OVER and UNDER range status indicators Analog outputs DC 10mV per dB AC 2Vrms full scale Pow...

Page 12: ...al or consequential damages FLIR s total liability is limited to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is express...

Reviews: