Specifications
General Specifications
Display
Multi-function
LCD
Transducers
30, 150 and 300 max. psi (sold separately)
Transducer signal
Full scale: 100mVDC
Display units
Bar, psi, Kg/cm
2
, mmHg, inches Hg, meters H
2
O, inches H
2
O,
Atmosphere
Zero adjust
Front panel push-button
Data Hold
Displayed reading can be held by front panel push-button
Min/Max Memory
Lowest/Highest readings can be stored for later recall
PC Interface kit
Built-in RS-232 port for use with optional Windows
TM
data
acquisition software
Auto Power-Off
To conserve battery life, the meter turns OFF automatically
after approx. 10 minutes (programmable over-ride).
Sampling time
0.8 seconds approx.
Power supply
9V Battery
Low battery indication
Rectangular battery-life box appears on the left side of LCD.
Power consumption
Approx. 7mA
Operating Temperature
32 to 122
o
F (0 to 50
o
C)
Operating Humidity
Less than 80% RH
Dimensions
7.1 x 2.8 x 1.3" (180 x 72 x 32mm)
Weight
Approx. 0.76 lbs. (345g)
Accessories
Protective rubber holster and 9V battery
Transducer Range, Resolution, and Accuracy Specifications
psi bar Kg/cm
2
mm Hg
in Hg
m H
2
O in
H
2
O atm
30 2
2.040 1500 59.05
20.4 802 1.974
Range
Resolution
0.02 0.002 0.002
2
0.05 0.02
1
0.002
150 10
10.19 7500 295.2 101.9 4010 9.87
Range
Resolution
0.2 0.01 0.01
10
0.2 0.1
5
0.01
300 20
20.40 15000 590.5 204.0 8020 19.74
Range
Resolution
0.2 0.02 0.02
20
0.5 0.2
10
0.02
Accuracy:
Meter only: ± (0.5% r 1 digit)
Transducer only: ± 2% full scale
Combined: ± (2.5% full scale + 1 digit)
Accuracy note:
Specified accuracy applies to meter at 23
o
C (±5
o
C)
Calibration note: Re-calibration is not required each time a new transducer is attached.