407495  V2.1   11/2007 

Operation 

Push-button Zero 

1.  Attach a transducer to the meter, turn the meter ON, and select the range that matches 

the attached transducer as described in the 'Preparation for Use' section. 

2.  With the transducer attached to the meter (but not attached to the process) press the 

ZERO button. Note that the ZERO button should always be pressed after swapping 
transducers. 

3.  If the display reads zero after pressing the ZERO button proceed to the next section. If 

the display does not zero, the transducer may be defective. 

Connect Transducer Port to Process 

1.  Ensure that the process under test is not pressurized 

2.  Connect the Transducer's ¼” PS, 19 teeth per inch threaded port to the process under 

test. 

3.  Apply pressure to the process. 

4.  Read the pressure on the meter's LCD. 

Data Hold 

1.  When a reading is on the display that you wish to hold, press the Data Hold button. 

2.  The reading will freeze and the HOLD icon will appear on the display. 

3.  Press the Data Hold button again to release the reading and return the meter to normal 

operation. 

Minimum/Maximum Reading Memory 

1.  Press the RECORD/RECALL button to start tracking the highest (MAX) and lowest 

(MIN) readings. The REC indicator will appear on the LCD. 

2.  Take pressure measurements. 

3.  When pressure measurements are complete, press the RECORD/RECALL button to 

recall the Maximum reading recorded (the MAX indicator will appear on the LCD). 

4.  Press the RECORD/RECALL button again to recall the Minimum reading recorded (the 

MIN indicator will appear on the LCD). 

5.  To exit this mode and return the meter to normal operation, press and hold the 

RECORD/RECALL button until the REC indicator goes out. 

Auto Power-Off 

In order to conserve battery life, the meter automatically shuts off after 10 minutes. To 
defeat this feature put the meter in the Record mode by pressing the RECORD/RECALL 
button; the REC indicator will appear. As long as the REC icon is displayed, the meter will 
stay on indefinitely. 

RS-232 PC Interface 

This meter is equipped with an RS-232 PC Interface output port (meter top). A Data 
Acquisition kit that includes Windows

TM

 95 / 98 / ME / NT / 2000 compatible software and a 

meter-to-PC interface cable can be obtained from Extech for use with this meter. The Part 
Number for the kit is 407001A. Instructions for use are supplied with the kit. 

Summary of Contents for 407495

Page 1: ...User s Manual Heavy Duty Pressure Meter Model 407495 ...

Page 2: ...0 for authorization or visit our website www extech com for contact information A Return Authorization RA number must be issued before any product is returned to Extech The sender is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation o...

Page 3: ... buttons when the LCD display switches on 3 A number will appear on the LCD that represents the Transducer type Select the Transducer type by pressing the SENSOR TYPE key until the desired transducer number appears Select the transducer from the list below in step 5 4 Press the ZERO button to store the new transducer type The meter will memorize the new type even when the meter is turned OFF 5 Sel...

Page 4: ...n Minimum Maximum Reading Memory 1 Press the RECORD RECALL button to start tracking the highest MAX and lowest MIN readings The REC indicator will appear on the LCD 2 Take pressure measurements 3 When pressure measurements are complete press the RECORD RECALL button to recall the Maximum reading recorded the MAX indicator will appear on the LCD 4 Press the RECORD RECALL button again to recall the ...

Page 5: ... Rectangular battery life box appears on the left side of LCD Power consumption Approx 7mA Operating Temperature 32 to 122 o F 0 to 50 o C Operating Humidity Less than 80 RH Dimensions 7 1 x 2 8 x 1 3 180 x 72 x 32mm Weight Approx 0 76 lbs 345g Accessories Protective rubber holster and 9V battery Transducer Range Resolution and Accuracy Specifications psi bar Kg cm 2 mm Hg in Hg m H2O in H2O atm 3...

Page 6: ...C º 0 1 e l a c S l l u F 2 o r e Z Operating Temperature 32ºF to 140ºF 0ºC to 40ºC Operating Humidity Max 80 RH Output Full capacity output 100 mV DC m m 5 8 x r e t e m a i d m m 0 3 e z i S Weight 5 65oz 160g Cable length 39 1m Calibration and Repair Services Extech offers repair and calibration services for the products we sell Extech also provides NIST certification for most products Call the...

Reviews: