407790A-en-US_v2.0 8/16
17
3.5
Selecting
the
Lmax/Lmin
type
Band
‐
(Band
maximum/band
minimum):
The
analysis
result
applies
to
the
point
where
the
level
for
each
frequency
band
was
maximum
or
minimum
within
the
sampling
period.
AP
‐
(All
‐
pass
maximum/all
‐
pass
minimum):
The
analysis
result
applies
to
the
sub
channel
all
‐
pass
and
band
‐
pass
level
values
taken
at
the
point
where
the
all
‐
pass
level
in
the
main
channel
was
maximum/minimum
within
the
sampling
period.
AP
(S)
‐
(All
‐
pass
maximum/all
‐
pass
minimum):
The
analysis
result
applies
to
the
band
‐
pass
level
values
(excluding
the
sub
channel
all
‐
pass
levels)
taken
at
the
point
where
the
all
‐
pass
level
in
the
main
channel
was
maximum/minimum
within
the
sampling
period.
The
sub
channel
all
‐
pass
level
is
taken
at
the
maximum/minimum
point
within
the
sampling
period,
independently
of
the
main
channel
level.
1.
Press
the
MENU
button
to
enter
the
menu
list
screen.
2.
Use
the
+
/
-
buttons
to
select
“
I).
Lmax/Lmin
Type
”
then
press
button.
3.
Use
the
+
/
-
buttons
to
select
the
desired
Band,
AP
or
AP
(S)
type
then
press
button.
4.
Press
MENU
button
to
return
to
the
measurement
screen.
This
setting
is
effective
in
SLM
mode
and
Octave
band
analyzer
mode.