380375 V1.5 3/04
3
Specifications
General specifications
Display
2 x 16 character alphanumeric multi-function dot matrix
Range selection
Automatic ranging
Bargraph
Displays voltage ramp, soak, and decay
Automatic discharge
After automatic & manual stop or upon completion of test
Low battery indicator
"Replace battery" displayed when battery voltage is low
Power source
Eight 1.5V AA cells; Battery life 40 hrs (no load at 5kV)
Auto Power off
After 30 minutes
Enersave
TM
mode
Shorter test time to conserve battery life
Operating conditions
32 to 104
o
F (0 to 40
o
C); < 80% RH
Storage conditions
14 to 140
o
F (-10 to 60
o
C); < 80% RH
Dimensions
4.7 x 6.7 x 3.7” (120 x 170 x 95mm)
Weight
1.94 lbs (880g)
Range Specifications
Test
Voltage
Resistance
Ranges
Short
Circuit
Current
Accuracy Resolution
500V DC
0 to 25 G
Ω
1mA
1000V DC
0 to 50 G
Ω
1mA
2500V DC
0 to 125 G
Ω
1mA
5000V DC
0 to 250 G
Ω
1.6 mA
0 to 50G
Ω
: ± 3% reading
50 to 250G
Ω
: ± 5% reading
1k
Ω
Meter Description
1. Test lead jacks compartment
2. Dot Matrix Display
3. Voltage test buttons
4. ON / TEST button
5. Battery compartment is
located on bottom of unit
1
\
2
3
4
5