SPECIFICATIONS

 

 

Ranges

 

Resolution

 

Accuracy

 

pH

 

0.00

 

to

 

14.00pH

 

0.01pH

 

±

 

0.02pH

 

Conductivity

 

0.0

 

to

 

200.0µS

 

200

 

to

 

2000µS

 

2.00

 

to

 

20.00mS

 

0.1µS

 

1µS

 

0.01mS

 

±

 

2%

 

FS

 

TDS

 

0.0

 

to

 

134.0ppm

 

134

 

to

 

1340ppm

 

1.34

 

to

 

13.40ppt

 

0.1ppm

 

1ppm

 

0.01ppt

 

(calculated

 

from

 

Conductivity)

 

Salinity

 

0.0

 

to

 

100.0ppm

 

100

 

to

 

1000ppm

 

1.00

 

to

 

10.00ppt

 

0.1ppm

 

1ppm

 

0.01ppt

 

(calculated

 

from

 

Conductivity)

 

ORP

 

1500

 

to

 

1500mV

 

1mV

 

±

 

3mV

 

Temperature

 

32

 

to

 

194ºF

 

0.0

 

to

 

90.0ºC

 

0.1º

 ≤ 

99.9º

 

 ≥ 

100º

 

±

 

2ºF/1ºC

 

(meter

 

+

 

probe)

 

 

Display

 

9999

 

count

 

LCD

  

MTC

 

temperature

 

range

 

32.0

 

to

 

194.0ºF

 

(0.0

 

to

 

90.0ºC)

 

pH

 

calibration

 

points

 

4.00,

 

7.00

 

and

 

10.00pH

 

Conductivity

 

calibration

 

points

 

84.0µS,

 

1413µS,

 

and

 

12.88mS

 

TDS

 

conversion

 

ratio

 

0.67

 

(fixed)

 

Salinity

 

conversion

 

ratio

 

0.5

 

(fixed)

 

Auto

 

Power

 

OFF

 

After

 

10

 

minutes

 

of

 

inactivity

 

(can

 

be

 

disabled)

 

Over

 

range

 

indication

  

“OL”

 

Operating

 

Temperature

 

41ºF

 

to

 

104ºF

 

(5ºC

 

to

 

40ºC)

 

Storage

 

Temperature

 

4ºF

 

to

 

140ºF

 

(

20ºC

 

to

 

60ºC)

 

Operating

 

Humidity

 

 

Max

 

80%

 

up

 

to

 

87ºF

 

(31ºC)

 

decreasing

 

linearly

 

to

 

50%

 

at

 

104ºF

 

(40ºC)

 

Storage

 

Humidity

 

<80%

  

Operating

 

Altitude

 

7000’

 

(2000m)

 

maximum

 

Power

 

9V

 

alkaline

 

battery

 

or

 

AC

 

adapter

 

Dimensions

 

4.7

 

x

 

3.8

 

x

 

1.8”

 

(118

 

x

 

96

 

x

 

45mm)

 

with

 

meter

 

closed

 

Weight

 

12oz.

 

(340g)

 

 
 

 

[email protected]

www.

GlobalScientificSupply

.com

GLOBAL SCIENTIFIC SUPPLY

The Laboratory Supply Company!

Summary of Contents for 341350A

Page 1: ...USER GUIDE Oyster Meter pH Conductivity TDS Salinity and ORP mV Model 341350A info GlobalScientificSupply com www GlobalScientificSupply com GLOBAL SCIENTIFIC SUPPLY The Laboratory Supply Company ...

Page 2: ...er Guide Product Updates Product Registration and Customer Support Meter Description Front and Side panels 1 LCD display 2 Battery compartment 3 Power button 4 Up Down arrow buttons 5 MODE CAL button 6 UNIT SET button 7 STORE RECALL button 8 Connector Conductivity probe 9 Connector pH probe ORP probe 10 Connector Temperature probe 11 Connector AC adaptor Display 1 Mode indicators 2 Primary display...

Page 3: ... and store it in the wetting cap with 4 0 M KCL or in a pH 4 buffer solution Conductivity TDS Salt Measurements 1 Press the POWER button to turn the meter ON 2 Press the MODE CAL button until the COND SALT or TDS icon is displayed 3 Connect the conductivity electrode to the side conductivity connector 4 If required calibrate the electrode see Conductivity Calibration 5 Place the electrode in the s...

Page 4: ...F or ºC Manual Temperature Compensation In the pH mode the solution temperature can be measured and set by using the optional external temperature probe or when the probe is not used by adjusting the temperature display Press the buttons to set the temperature when the MTC icon is displayed The Conductivity probe has a built in temperature sensor so manual temperature compensation does not apply f...

Page 5: ...calibrates itself to that value At the end of the calibration cycle SA and End will briefly appear on the display and the meter will return to the normal operating mode 5 When a calibration is performed the calibration icons pH10 pH7 pH4 will be cleared calibration data is not erased and will be replaced when a successful calibration is performed for each buffer within one power on cycle These ico...

Page 6: ...the end of the calibration the display will briefly indicate SA End and then return to the measurement mode Note The SA will not appear if the calibration fails 5 The range calibrated alert will appear in the display for each range that is calibrated during that power on cycle Low range 84µS cm Medium range 1413µS cm High range 12 88mS cm 12 880µS cm Note The meter allows for a 1 2 or 3 point cali...

Page 7: ...solution 2 If bubbles are seen in the bulb area hold the electrode by its cap and shake downwards until bubbles are removed 3 To improve speed of response vigorously stir the electrode in the sample buffer or rinse solution 4 After exposure to a sample buffer or rinse solution shake the electrode with a snap motion to remove residual drops of solution 5 When possible use part of the next sample bu...

Page 8: ...94 0ºF 0 0 to 90 0ºC pH calibration points 4 00 7 00 and 10 00pH Conductivity calibration points 84 0µS 1413µS and 12 88mS TDS conversion ratio 0 67 fixed Salinity conversion ratio 0 5 fixed Auto Power OFF After 10 minutes of inactivity can be disabled Over range indication OL Operating Temperature 41ºF to 104ºF 5ºC to 40ºC Storage Temperature 4ºF to 140ºF 20ºC to 60ºC Operating Humidity Max 80 up...

Page 9: ... to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied Calibration Repair and Customer Care Services FLIR Systems Inc offers repair and calibration services for the Extech Instruments products we sell NIST certification for most products is also provided Call the Customer Service Department for inform...

Page 10: ...paration ou au remplacement du produit La garantie définie ci dessus est inclusive et aucune autre garantie écrite ou orale n est exprimée ou implicite Calibrage réparation et services après vente FLIR Systems Inc offre des services de calibrage et de réparation pour les produits Extech Instruments que nous commercialisons Nous fournissons également une certification NIST pour la plupart des produ...

Page 11: ... está limitada a la reparación o reemplazo del producto La garantía precedente es inclusiva y no hay otra garantía ya sea escrita u oral expresa o implícita Servicios de calibración reparación y atención a clientes FLIR Systems Inc ofrece servicios de reparación y calibración para los productos que vendemos de Extech Instruments Además ofrecemos certificación NIST para la mayoría de los productos ...

Reviews: