Extech Instruments 341350A User Manual Download Page 7

341350A-en-US_V1.3   8/15 

7

pH

 

Troubleshooting

 

Chart

 

Symptom

 

Cause

 

Recommended

 

Solution

 

Long

 

response

 

time

 

or

 

reading

 

drift

 

Clogged

 

Junction

 

Soak

 

in

 

4.07

 

M

 

KCL

 

@

 

60

o

C

 

for

 

30

 

minutes

 

Oil,

 

paint,

 

dyes,

 

suspended

 

solids

 

on

 

sensor

 

Rinse

 

electrode

 

alternately

 

with

 

materials

 

solvent

 

then

 

buffer

 

7.00

 

Dry

 

Bulb

 

Long

 

term

 

storage

 

without

 

wetting

  

Soak

 

electrode

 

tip

 

in

 

wetting

 

cap

 

filled

 

with

 

1ml

 

7.00

 

buffer

 

for

 

24

 

to

 

48

 

hours

 

Static

 

Charge

 

Wiping

 

electrodes

 

Rinse

 

electrode

 

in

 

7.0

 

buffer

 

and

 

blot.

 

Do

 

no

 

wipe

 

electrode.

 

Same

 

readings

 

in

 

different

 

buffers

 

and

 

samples

 

Cracked

 

or

 

broken

 

bulb

 

Replace

 

electrode.

 

Use

 

bulb

 

guard.

 

Avoid

 

plunging

 

electrode

 

to

 

bottom

 

of

 

container

 

and

 

spinning

 

bars.

 

A

 

wetting

 

cap

 

will

 

protect

 

bulb

 

between

 

measurements.

 

Erratic

 

LCD

 

display

 

Samples

 

have

 

low

 

ionic

 

strength

 

(lacks

 

salt);

 

e.g.

 

distilled,

 

de

ionized,

 

boiled,

 

lake

 

water

 

(high

 

pressure)

 

For

 

each

 

50

 

ml

 

of

 

sample

 

add

 

1

 

drop

 

(50uL)

 

of

 

SAT.KCL

 

No

 

alteration

 

in

 

pH

 

will

 

occur

 

by

 

inert

 

KCL.

 

 

Notes

 

on

 

pH

 

measurements

 

and

 

electrodes

 

1. 

The

 

Electrode

 

should

 

be

 

stored

 

in

 

its

 

wetting

 

cap

 

until

 

used.

 

Use

 

a

 

pH

 

4

 

buffer

 

solution

 

or

 

a

 

4.0

 

M

 

KCL

 

solution.

 

2. 

If

 

bubbles

 

are

 

seen

 

in

 

the

 

bulb

 

area,

 

hold

 

the

 

electrode

 

by

 

its

 

cap

 

and

 

shake

 

downwards

 

until

 

bubbles

 

are

 

removed.

 

3. 

To

 

improve

 

speed

 

of

 

response,

 

vigorously

 

stir

 

the

 

electrode

 

in

 

the

 

sample,

 

buffer,

 

or

 

rinse

 

solution.

 

4. 

After

 

exposure

 

to

 

a

 

sample,

 

buffer,

 

or

 

rinse

 

solution,

 

shake

 

the

 

electrode

 

with

 

a

 

snap

 

motion

 

to

 

remove

 

residual

 

drops

 

of

 

solution.

 

5. 

When

 

possible,

 

use

 

part

 

of

 

the

 

next

 

sample/buffer

 

to

 

be

 

measured

 

as

 

a

 

rinse

 

solution,.

 

6. 

Keep

 

buffers

 

and

 

samples

 

at

 

the

 

same

 

temperature

 

to

 

avoid

 

temperature

 

effects.

 

7. 

pH

 

readings

 

stabilize

 

faster

 

in

 

some

 

solutions

 

than

 

others;

 

allow

 

time

 

to

 

stabilize.

 

8. 

Electrodes

 

deteriorate

 

over

 

time.

 

If

 

accuracy

 

falls

 

to

 

10%

 

the

 

electrode

 

should

 

be

 

cleaned.

 

If

 

no

 

improvement

 

is

 

observed,

 

replace

 

the

 

electrode.

 

Notes

 

on

 

the

 

Conductivity/TDS/Salt

 

Cell

 

1. 

Cell

 

Storage:

 

On

 

sheathed

 

cells,

 

replace

 

the

 

sheath

 

over

 

the

 

cell

 

when

 

storing.

 

For

 

non

sheathed

 

versions,

 

soak

 

the

 

cell

 

tip

 

in

 

de

ionized

 

water

 

for

 

storage.

 

2. 

Cell

 

Cleaning:

 

After

 

each

 

use,

 

the

 

cell

 

tip

 

should

 

be

 

rinsed

 

with

 

de

ionized

 

water.

 

If

 

solids

 

build

 

up

 

inside

 

the

 

cell

 

carefully

 

remove

 

with

 

a

 

cotton

 

swab

 

soaked

 

in

 

solvent

 

taking

 

care

 

not

 

to

 

touch

 

the

 

metal

 

parts

 

of

 

the

 

inner

 

cell.

 

[email protected]

www.

GlobalScientificSupply

.com

GLOBAL SCIENTIFIC SUPPLY

The Laboratory Supply Company!

Summary of Contents for 341350A

Page 1: ...USER GUIDE Oyster Meter pH Conductivity TDS Salinity and ORP mV Model 341350A info GlobalScientificSupply com www GlobalScientificSupply com GLOBAL SCIENTIFIC SUPPLY The Laboratory Supply Company ...

Page 2: ...er Guide Product Updates Product Registration and Customer Support Meter Description Front and Side panels 1 LCD display 2 Battery compartment 3 Power button 4 Up Down arrow buttons 5 MODE CAL button 6 UNIT SET button 7 STORE RECALL button 8 Connector Conductivity probe 9 Connector pH probe ORP probe 10 Connector Temperature probe 11 Connector AC adaptor Display 1 Mode indicators 2 Primary display...

Page 3: ... and store it in the wetting cap with 4 0 M KCL or in a pH 4 buffer solution Conductivity TDS Salt Measurements 1 Press the POWER button to turn the meter ON 2 Press the MODE CAL button until the COND SALT or TDS icon is displayed 3 Connect the conductivity electrode to the side conductivity connector 4 If required calibrate the electrode see Conductivity Calibration 5 Place the electrode in the s...

Page 4: ...F or ºC Manual Temperature Compensation In the pH mode the solution temperature can be measured and set by using the optional external temperature probe or when the probe is not used by adjusting the temperature display Press the buttons to set the temperature when the MTC icon is displayed The Conductivity probe has a built in temperature sensor so manual temperature compensation does not apply f...

Page 5: ...calibrates itself to that value At the end of the calibration cycle SA and End will briefly appear on the display and the meter will return to the normal operating mode 5 When a calibration is performed the calibration icons pH10 pH7 pH4 will be cleared calibration data is not erased and will be replaced when a successful calibration is performed for each buffer within one power on cycle These ico...

Page 6: ...the end of the calibration the display will briefly indicate SA End and then return to the measurement mode Note The SA will not appear if the calibration fails 5 The range calibrated alert will appear in the display for each range that is calibrated during that power on cycle Low range 84µS cm Medium range 1413µS cm High range 12 88mS cm 12 880µS cm Note The meter allows for a 1 2 or 3 point cali...

Page 7: ...solution 2 If bubbles are seen in the bulb area hold the electrode by its cap and shake downwards until bubbles are removed 3 To improve speed of response vigorously stir the electrode in the sample buffer or rinse solution 4 After exposure to a sample buffer or rinse solution shake the electrode with a snap motion to remove residual drops of solution 5 When possible use part of the next sample bu...

Page 8: ...94 0ºF 0 0 to 90 0ºC pH calibration points 4 00 7 00 and 10 00pH Conductivity calibration points 84 0µS 1413µS and 12 88mS TDS conversion ratio 0 67 fixed Salinity conversion ratio 0 5 fixed Auto Power OFF After 10 minutes of inactivity can be disabled Over range indication OL Operating Temperature 41ºF to 104ºF 5ºC to 40ºC Storage Temperature 4ºF to 140ºF 20ºC to 60ºC Operating Humidity Max 80 up...

Page 9: ... to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied Calibration Repair and Customer Care Services FLIR Systems Inc offers repair and calibration services for the Extech Instruments products we sell NIST certification for most products is also provided Call the Customer Service Department for inform...

Page 10: ...paration ou au remplacement du produit La garantie définie ci dessus est inclusive et aucune autre garantie écrite ou orale n est exprimée ou implicite Calibrage réparation et services après vente FLIR Systems Inc offre des services de calibrage et de réparation pour les produits Extech Instruments que nous commercialisons Nous fournissons également une certification NIST pour la plupart des produ...

Page 11: ... está limitada a la reparación o reemplazo del producto La garantía precedente es inclusiva y no hay otra garantía ya sea escrita u oral expresa o implícita Servicios de calibración reparación y atención a clientes FLIR Systems Inc ofrece servicios de reparación y calibración para los productos que vendemos de Extech Instruments Además ofrecemos certificación NIST para la mayoría de los productos ...

Reviews: