Intended use

The intended use of EXOPULSE Mollii Suit is: 

- Relaxation of spastic muscles

- Maintaining or increasing range of motion

- Activation and re-education of muscles

- Retardation or prevention of disuse atrophy 

- Increasing local blood circulation 

- Symptomatic relief and management of chronic intractable pain

- Adjunctive treatment for post-surgical pain

EXOPULSE Mollii Suit is for use in the home environment.

Change of name

EXOPULSE Mollii Suit was previously called Mollii.

EXOPULSE Mollii Suit consists of

EXOPULSE Body Garments (item number: 100-1)

1 jacket and 1 pair of trousers supplied with laundry 

bags.

EXOPULSE Control Unit (item number: 100-2)

1 control unit. The control unit can be connected to the 

garment unit by means of 4 magnetic couplings. The 

control unit is secured around the waist with the belt 

supplied.

Item number

Note that every size of jacket and trousers has a unique item number. For a complete list 

of item numbers see: www.exopulse.com

EN

Summary of Contents for Mollii Suit

Page 1: ...Please read through the entire user manual prior to use User Manual EXOPULSE Mollii Suit...

Page 2: ......

Page 3: ...produced by EXONEURAL NETWORK and applies to EXOPULSE Mollii Suit The operating instructions are valid from 31 08 2020 Software version 1 3x EXONEURAL NETWORK AB Sv rdv gen 11D SE 182 33 Danderyd Swed...

Page 4: ...4...

Page 5: ...5 ENGLISH...

Page 6: ...or use in the home environment Change of name EXOPULSE Mollii Suit was previously called Mollii EXOPULSE Mollii Suit consists of EXOPULSE Body Garments item number 100 1 1 jacket and 1 pair of trouser...

Page 7: ...is in PAUSE mode Symbol Dry flat Display Product defect do not use the product Contact EXONEURAL NETWORK AB if this symbol is displayed Symbol Do not dry clean Display Battery indicator Symbol Do not...

Page 8: ...the above can expose the user to unnecessary risk and is therefore at the user s own risk Always consult a doctor EXOPULSE Mollii Suit is for use solely by the user for whom it is adapted Magnetic cou...

Page 9: ...agnetic arms should be tested for natural tear and wear once per year and replaced if faulty The replacement must only be done by an approved distributor and will be covered by the product warranty pr...

Page 10: ...ipes for this as they are gentle on the skin REGULARLY The garment can be machine washed in a laundry bag at 40 C delicate wash Wash regularly about once a month to keep fabric and electrodes clean Cl...

Page 11: ...ect the magnetic couplings in accordance with the picture on the right See the next page for dressing instructions The electrodes end up in the correct position if the right size is used and the zips...

Page 12: ...sing 1 Start with the trousers Ensure that the lower zip ends up on the outside of the leg and forms a line between the ankle and the outside of the knee 2 Check that the zips on the trousers form a s...

Page 13: ...trousers Do up the zip on the stomach first then the zips for the arms Check that the garment unit is positioned correctly on the body and the zips are straight 6 Thread the belt through the loops on...

Page 14: ...to continue FINISH Turn off the control unit using the ON OFF button Remove the control unit before taking off the jacket and trousers Clean and store EXOPULSE Mollii Suit according to the instructio...

Page 15: ...is prescribed Depending on diagnosis and manual examination of the patient the prescriber decides which muscles are to be stimulated and with which pulse width The user is not able to set pulse pa ram...

Page 16: ...ther electronic equipment Guidance and the manufacturer s declaration electromagnetic radiation EXOPULSE Mollii Suit is intended to be used in an electromagnetic environment as below It is the custome...

Page 17: ...es 1 kV for input output lines N A Not relevant in this case EXOPULSE Mollii Suit is oper ated by an internal voltage source bat teries Overvoltage protection IEC 61000 4 5 1 kV differential mode 2 kV...

Page 18: ...uency interval b Interference can arise in the vicinity of equipment marked with the symbol below NOTE 1 The higher frequency interval applies at 80MHz and 800MHz NOTE 2 This guidance does not apply i...

Page 19: ...output voltage of the trans mitter W The distance depends on the frequency of the transmitter m 150 kHz to 80 MHz P d 2 1 80 MHz to 800 MHz P d 2 1 800 MHz to 2 5 GHz P d 3 2 0 01 0 12 0 12 0 24 0 1...

Page 20: ...he event of accidents or incidents Use the contact details below EXONEURAL NETWORK Sv rdv gen 11D 182 33 Danderyd Sweden Tel 46 0 8 410 277 01 Mail support exopulse com If equipment is defective conta...

Page 21: ...21...

Page 22: ...EXONEURAL NETWORK AB Sv rdv gen 11D SE 182 33 Danderyd Sweden Tel 46 0 8 410 277 01...

Reviews: