18
Guidance and the manufacturer’s certification – electromagnetic immunity
EXOPULSE Mollii Suit is intended to be used in an electromagnetic environment as below. It is the cus-
tomer’s and the user’s responsibility to ensure that usage is limited to the environment described.
Immunity test
IEC 60601 test level
Compliance level Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80MHz to 2,5GHz
10Vrms
10 V/m
Portable- and mobile RF communication
equipment with accompanying leads
should be kept at the recommended
distance from EXOPULSE Mollii Suit.
The distance can be calculated using the
equation below and the frequency of
radiation from the radiation source.
Recommended distance:
150 kHz - 80 MHz:
P
d
2
,1
=
80 MHz - 2.5 GHz:
P
d
3
,
2
=
Where P is the maximum outgoing
output of the transmitter in watts (W)
from the manufacturer of the equipment
and d is the recommended distance in
metres (m).
The field strength from fixed radio fre-
quency radioactive sources is determined
by an electromagnetic site investigation
a), but should be lower than the level for
the respective frequency interval b).
Interference can arise in the vicinity
of equipment marked with the symbol
below:
NOTE 1 The higher frequency interval applies at 80MHz and 800MHz.
NOTE 2 This guidance does not apply in all situations and environments. The electromagnetic environ-
ment can be affected by absorption and reflection from objects, people and other environmental factors.
a) Field strengths from fixed transmitters, such as base stations for radio (mobile or wireless) telephones,
mobile radio, amateur radio, AM and FM transmissions and TV broadcasts cannot be predicted with
theoretical accuracy. A site investigation should be considered to evaluate the electromagnetic environ-
ment on the basis of these factors. If the field strength measured in locations where EXOPULSE Mollii
Suit EXOPULSE Mollii Suit is going to be used exceeds the above RF values, use of EXOPULSE Mollii
Suit EXOPULSE Mollii Suit should be monitored and regulated so that normal function can be ensured. If
variations in functionality are observed, further measurements, as well as moving and testing EXOPULSE
Mollii Suit EXOPULSE Mollii Suit in other environments, should be implemented.
b) The field strength for the 150 kHz -80MHz frequency interval must be less than 10 V/m.
Recommended distance between portable and mobile RF communication equipment and EXOPULSE
Mollii Suit
EN