Remote Control Commands
22
T100S-HP Programming Guide
2.4
Diode Current Settings
2.4.1
I
Syntax
I=xxx.x
Parameter
•
xxx.x
: current level in mA, in the range 0-400 mA.
Description
Sets the laser current level (in mA) and switches to the constant-current mode (see
section
).
Response
•
RS-232C:
OK
•
GPIB: none, see section
Checking System Status with the Status Byte, p. 9
2.4.2
I?
Syntax
I?
Parameter
None.
Description
Returns the value of the diode current in mA.
The module optical output must be enabled (see section
).
Response
•
I=xxx.x
: diode value in mA.
•
DISABLED
: the optical output is disabled, the current level value cannot be returned.
2.4.3
LIMIT?
Syntax
LIMIT?
Parameter
None.
Description
Returns the status of the laser diode current limitation (400 mA).
Response
•
YES
: the current has reached its limit.
•
NO
: the current has not reached its limit.