Remote Control Commands
T100S-HP Programming Guide
21
Description
Sets mW as the unit for optical power values.
Response
•
RS-232C:
OK
•
GPIB: none, see section
Checking System Status with the Status Byte, p. 9
2.3.3
P
Syntax
P=[±]xx.xx|xx.xx
Parameter
•
[±]xx.xx
: optical output power in dBm, if the unit is set to dBm (see section
). Possible values are given in the
Technical Specifications
section in
T100S-HP
User Manual
.
•
xx.xx
: optical output power in mW, if the unit is set to mW (see section
).
Possible values are given in the
Technical Specifications
section in
T100S-HP User
Manual
.
Description
Sets the optical power (in dBm or mW depending on the selected power unit) and
switches to the constant-power mode (see section
).
The units (dB or mW) must be first selected with the appropriate DBM or MW command.
Response
•
RS-232C:
OK
•
GPIB: none, see section
Checking System Status with the Status Byte, p. 9
2.3.4
P?
Syntax
P?
Parameter
None.
Description
Returns the current value of the laser output power, according to the selected power unit.
The format of the response depends on the power unit selected (see section
Response
•
P=xx.xx
: output power value in mW.
•
P=±xx.xx
: output power value in dBm.
•
DISABLED
: the optical output is disabled, the output power value cannot be returned.