Maintenance
54
IQS-9100
Cleaning EUI Connectors
Cleaning EUI Connectors
Regular cleaning of EUI connectors will help maintain optimum
performance. There is no need to disassemble the unit.
To clean EUI connectors:
1.
Remove the EUI from the instrument to expose the connector
baseplate and ferrule.
2.
Moisten a 2.5 mm cleaning tip with
one drop
of isopropyl alcohol
(alcohol may leave traces if used abundantly).
3.
Slowly insert the cleaning tip into the EUI adapter until it comes out on
the other side (a slow clockwise rotating movement may help).
4.
Gently turn the cleaning tip one full turn, then continue to turn as you
withdraw it.
I
MPORTANT
If any damage occurs to internal connectors, the module casing will
have to be opened and a new calibration will be required.
Push
Turn
Pull
3
4
5