Introducing the FIP-500 Fiber Inspection Scope
6
FIP-500
LED Indicators Description
Dual-Function LED
The dual-function LED is located just above the trigger used for captures,
on the front of your unit. It can indicate the status of the inspection or be
used as a flashlight to help you see the connectors you have to inspect in
darker environments (see
Turning on or Turning off the Flashlight
The table below indicates the possible statuses.
Status
Meaning
Blue
The unit is ready and operational.
The flashlight feature is disabled.
Blue, blinking
The capture process is in progress.
Green, steady for 5 seconds
The analysis found no defects or scratches exceeding
the pass criteria.
Red, blinking for 5 seconds
The capture process could not be completed
successfully. Follow the instructions on the screen.
The auto-focus operation has timed out.
There is a mismatch between the number of fibers
found and the selected multi-fiber connector layout.
Red, steady for 5 seconds
The analysis found defects or scratches that do not
meet the pass criteria.
White
The unit is ready and operational.
The flashlight feature is enabled.
Summary of Contents for FIP-500
Page 1: ...User Guide Fiber Inspection Scope FIP 500 www EXFO com...
Page 10: ......
Page 20: ......
Page 50: ......
Page 62: ......
Page 81: ...Maintenance Fiber Inspection Scope 71 Cleaning Lenses Lens Single fiber optical head...
Page 98: ...7 Pull the battery up to remove it Socket for battery connection...
Page 124: ......
Page 130: ...0 5 1 5 48 5 0 176 W d y K y K D...