background image

5

Thermal Scanning

The high-performance Microscanner

 D-Series can also be used to  

measure tempera tures at a distance.

When not touching or close to a surface:
•  The Microscanner

 is a wide-angle ther mal scanner with a fixed 

emissivity of .98.

•  The displayed reading is an average tem perature of everything that the 

sensor sees within its field of view.

Thermal scanning is very useful in many applications such as evaluating 
insulation, monitoring machinery, and inspecting electrical equipment.

The MAX or MIN modes of operation can be particularly useful in thermal 
scanning. The highest (or lowest) temperature measured dur ing a scan is 
indicated by the audible beeper. Once a hot spot (or cold spot) is identified, 
the exact temperature of that spot can be measured by touching the  
surface as previously ex plained.

When using the D-Series as a scanner, keep the following in mind:

•  The instrument’s field of view is 1:1. Also referred to as a distance-to-

spot ratio, 1:1 field of view means that the sensor sees a cir cular area 
with a diameter equal to the distance between the sensor and target.  
For example, at a 6-inch distance, the sensor sees a 6-inch diameter 
spot; at a 10-foot distance, the sensor sees a 10-foot diameter spot.

•  The Microscanner

 measures the average temperature of everything  

in its field of view. 

•  A small hot spot may get lost in a large viewing area. The closer you  

hold the instru ment to a surface, the better its target resolu tion.

Summary of Contents for Microscanner D501

Page 1: ...Instruction Manual Microscanner D Series D501 RS D501 High Performance Models...

Page 2: ...our high performance Microscanner D Series features Full range resolution to 0 1 F or C SCAN MAX and MIN modes of operation A 10 second display lock An audible beeper to signal functional or condition...

Page 3: ...initiate a new set of readings PUSH the power pushbutton at any time Operation Mode SCAN In the SCAN mode the target s instantaneous temperature is continuously displayed and updates 10 times per seco...

Page 4: ...vides a marker with each unit It is only necessary to mark an area slightly larger than your Microscanner s nosepiece In the MAX or MIN mode you may want to check several points over an area In this c...

Page 5: ...pment If electrical inspection is of particular importance contact Exergen for more informa tion relative to the dedicated electrical inspection instruments If the surface is metallic be sure to mark...

Page 6: ...nose piece Although this will not harm the unit the indicated temperature may be a few degrees too high If this happens allow a few minutes between readings for the nosepiece to cool down and stabili...

Page 7: ...a scan is indicated by the audible beeper Once a hot spot or cold spot is identified the exact temperature of that spot can be measured by touching the surface as previously ex plained When using the...

Page 8: ...s This assures that the water on the wet membrane is evaporating at its maximum rate 3 With the Microscanner in the MIN mode quickly measure the temperature of the WET membrane immediately after you s...

Page 9: ...p each second Temperatures will continue to be displayed as long as accuracy can be assured If the battery drops below 5 7 volts the battery is considered dead and the display defaults to If a problem...

Page 10: ...ar ent material The shiny metal will reflect the transparent materi al s heat back to the Microscanner Wet or icy surfaces may interfere with the Microscanner s readings Water and other liquids are go...

Page 11: ...nance required is to keep the sensor window clean This window located at the base of the nosepiece s reflective cup is made of a special crystal which passes infrared heat Dirt greasy films or moistur...

Page 12: ...sure If after applying all of the above solutions the instrument s accuracy is still in question contact Exergen for further directions Warranty Exergen warrants the Microscanner D Series to be free o...

Page 13: ...mm Repeatability 02 of absolute temperature Display Resolution 0 1 F or 0 1 C Display Update 10 times second Response Time Constant 60 mSec Field of View 1 1 53 degrees Spectral Sensitivity 2 to 20 mi...

Page 14: ...Sensor Window Model D501 Nosepiece with Reflective Cup AECS Mode Selector Switch On Off Power Pushbutton High impact ABS Case Digital LED Display Microscanner D Series Features Overview...

Page 15: ...Remote Sensor Probe Model D501 RS Microscanner D Series Features Overview...

Page 16: ...Exergen Corporation 400 Pleasant Street Watertown MA 02472 U S A 617 923 9900 www exergen com 818098 Rev 3 Microscanner D Series...

Reviews: