1

Operations and Controls

Your high-performance Microscanner

 D-Series is equipped with an ON/OFF 

power pushbutton and a mode selector switch. The mode selector switch allows 
you to choose one of three modes of operation: SCAN, MAX or MIN.

On/Off, Lock and Reset:

•  TO TURN the unit ON, PUSH and HOLD the ON/OFF power pushbutton. 

The display will momentarily read 8888, an indication that the micropro-
cessor is performing its self-diagnostic testing. After the test, the unit will 
measure and display temperature in the selected mode of operation for 
as long as the power button is depressed.

•  RELEASE the power pushbutton TO LOCK the display on the last reading. 

A single beep will audibly indicate that the display is locked. The instru-
ment will hold the last reading on the display for 10 seconds before it 
automatically turns itself OFF.

•  To RESET the display and initiate a new set of readings, PUSH the power 

pushbutton at any time.

Operation Mode:
•  SCAN

 - In the SCAN mode, the target’s instantaneous temperature is 

continuously displayed and updates 10 times per second for as long as 
you hold the power button down. After you release the power button, the 
display will lock on the last temperature measured and hold that reading 
for 10 seconds.

•  MAX

 - In the MAX mode, the display will lock on the highest temperature 

measured for as long as you hold the power button down. Each time a 
new peak temperature is measured or repeated an audible beep will 
sound. After you release the power button, the display will lock on the 
maximum recorded temperature and hold that reading for 10 seconds.

•  MIN

 - In the MIN mode, the display will lock on the lowest temperature 

measured for as long as you hold the power button down. Each time 
a new low temperature is measured or repeated, an audible beep will 
sound. After you release the power button, the display will lock on the 
minimum recorded temperature and hold that reading for 10 seconds.

Summary of Contents for Microscanner D501

Page 1: ...Instruction Manual Microscanner D Series D501 RS D501 High Performance Models...

Page 2: ...our high performance Microscanner D Series features Full range resolution to 0 1 F or C SCAN MAX and MIN modes of operation A 10 second display lock An audible beeper to signal functional or condition...

Page 3: ...initiate a new set of readings PUSH the power pushbutton at any time Operation Mode SCAN In the SCAN mode the target s instantaneous temperature is continuously displayed and updates 10 times per seco...

Page 4: ...vides a marker with each unit It is only necessary to mark an area slightly larger than your Microscanner s nosepiece In the MAX or MIN mode you may want to check several points over an area In this c...

Page 5: ...pment If electrical inspection is of particular importance contact Exergen for more informa tion relative to the dedicated electrical inspection instruments If the surface is metallic be sure to mark...

Page 6: ...nose piece Although this will not harm the unit the indicated temperature may be a few degrees too high If this happens allow a few minutes between readings for the nosepiece to cool down and stabili...

Page 7: ...a scan is indicated by the audible beeper Once a hot spot or cold spot is identified the exact temperature of that spot can be measured by touching the surface as previously ex plained When using the...

Page 8: ...s This assures that the water on the wet membrane is evaporating at its maximum rate 3 With the Microscanner in the MIN mode quickly measure the temperature of the WET membrane immediately after you s...

Page 9: ...p each second Temperatures will continue to be displayed as long as accuracy can be assured If the battery drops below 5 7 volts the battery is considered dead and the display defaults to If a problem...

Page 10: ...ar ent material The shiny metal will reflect the transparent materi al s heat back to the Microscanner Wet or icy surfaces may interfere with the Microscanner s readings Water and other liquids are go...

Page 11: ...nance required is to keep the sensor window clean This window located at the base of the nosepiece s reflective cup is made of a special crystal which passes infrared heat Dirt greasy films or moistur...

Page 12: ...sure If after applying all of the above solutions the instrument s accuracy is still in question contact Exergen for further directions Warranty Exergen warrants the Microscanner D Series to be free o...

Page 13: ...mm Repeatability 02 of absolute temperature Display Resolution 0 1 F or 0 1 C Display Update 10 times second Response Time Constant 60 mSec Field of View 1 1 53 degrees Spectral Sensitivity 2 to 20 mi...

Page 14: ...Sensor Window Model D501 Nosepiece with Reflective Cup AECS Mode Selector Switch On Off Power Pushbutton High impact ABS Case Digital LED Display Microscanner D Series Features Overview...

Page 15: ...Remote Sensor Probe Model D501 RS Microscanner D Series Features Overview...

Page 16: ...Exergen Corporation 400 Pleasant Street Watertown MA 02472 U S A 617 923 9900 www exergen com 818098 Rev 3 Microscanner D Series...