22
8.7. Other commands
POWER_OFF
Use this command to switch-off the device.
Byte0
Byte1
Byte2
Byte3
Byte4
0x32
0x32
0x32
0x32
CHKSUM
The sensor responds with an acknowledge, right before switching-off.
Byte0
0x01
This command is only accepted if the sensor is not in sampling/streaming mode.
RESTORE_PARAMETERS
Use this command to restore all settings to the default values.
Byte0
Byte1
0x67
0x67
The sensor responds with an acknowledge.
Byte0
0x01
This command is only accepted if the sensor is not in sampling/streaming mode.
NOTE: This command does not reset the
calibration parameters
START_BOOTLOADER
Use this command to start the
bootloader
at the next device’s switching-on.
Byte0
Byte1
Byte2
Byte4
0x80
0xBA
0xBA
CHKSUM
The device responds with an acknowledge..
Byte0
0x01
This command is only accepted if the sensor is not in sampling/streaming mode.
NOTE: This command is only supported by devices with
bootloader v.1.05
and above.