Care and Maintenance
The System Controllers are designed to provide years of trouble-free operation with little maintenance
activity. Here are some guidelines for ensuring long-life from the System Controller.
Do not locate the System Controller directly in an environment with air-borne contamination. If the
unit needs to be in such an environment, install inside a ventilated cabinet with fan-assisted air-
filtration. Observe the air flow direction of the SC-series (left to right) relative to the cabinet and
configure the cabinet accordingly.
Ensure all air-intakes and exhaust vents are at least 50mm away from obstruction and free from dust
and debris. If necessary, wipe with a dry cloth or use compressed air to dislocate particulate matter
which may be stuck to the vents.
Caution:
Do not use a wet cloth or any aerosol-type cleaning fluids as this may be ingested by the
unit, causing damage and the risk of electrical shock.
If the unit needs to be cleaned more thoroughly, first disconnect from the AC mains. Using a damp
cloth (impregnated with either with water or alcohol), wipe the surfaces of the unit. Take care to
ensure that the cloth is not dripping. If using alcohol, avoid the labels to prevent them from being
smeared or detaching from the unit and avoid heavy rubbing. Do not use acetone.
Ensure that AC-side and DC-side cables are connected firmly to the chassis and (where possible)
screwed in to prevent damage to the connectors. Cables should be tied off to provide strain relief to
the electrical connectors.
Do not use the electrical breaker as the means to turn on/off power on a regular basis. The switch on
the front of the unit is designed for more on-off cycles. The breaker should be set once when the
system is commissioned and only reset if it trips at a later date.