6
Dip the
eXact
®
Strip Micro CLH, Part No. 486672
into the
CELL
and immediately press
READ
.
This starts the
20 SECOND
countdown timer. During this time move the strip in a gentle back
and forth motion (approx. 2 strokes/sec).
Remove and discard the strip after “1” on the display
disappears
. The meter will begin counting up for
120 seconds
, at the end of which, the cursor will
move across the display while the meter prepares to measure the sample. Record result displayed
(this result is automatically stored in HR8). After testing is completed, rinse cell immediately.
*NOTE:
8VHDVHFRQGGLSWLPHLIZDWHUWHPSHUDWXUHLVDERYHÛ&)
DIP STRIP -
(read carefully and follow procedure closely)
5
Press the
ZERO/ON
button. The cursor will move across the display, followed by
0.00 PPM
.
The sample is ready for testing.
ZERO METER
2
Press the
ZERO/ON
button to power the meter on; the display will show all annunciators, then
the current MENU selection, followed by the last reading.
TURN METER ON
3
Press and re-press the
MENU
button until the display shows the parameter HR8.
SELECT TEST: HR8
4
Rinse the
CELL
at least 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\ZLWK
the water sample.
FILL METER WITH SAMPLE
9
Total Chlorine High Test Procedure
CLH
HR8
1
Remove one (1)
eXact
®
Strip Micro CLH, Part No. 486672
from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
1
Remove one (1)
eXact
®
Strip Micro
PH, Part No. 486639
from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
6
Dip the
eXact
®
Strip Micro PH, Part No. 486639
into the
CELL
and immediately press
READ
.
This starts the
20 SECOND
countdown timer. During this time move the strip in a gentle back
and forth motion (approx. 2 strokes/sec).
Remove and discard the strip after “1” on the display
disappears
. The cursor will move across the display while the meter prepares to measure the
sample. Record result displayed (this result is automatically stored in PH7). After testing is
completed, rinse cell immediately.
DIP STRIP AND PRESS “READ”
5
Press the
ZERO/ON
button. The cursor will move across the display, followed by
0.00 pH
.
The sample is ready for testing.
ZERO METER
2
Press the
ZERO/ON
button to power the meter on; the display will show all annunciators, then
the current MENU selection, followed by the last reading.
TURN METER ON
3
Press and re-press the
MENU
button until the display shows the parameter
PH7
.
SELECT TEST: PH7
4
Rinse the
CELL
at least 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\ZLWK
the water sample.
FILL METER WITH SAMPLE
pH Test Procedure
PH
PH7
NOTE: This test is good for all water temperatures 0°C - 40°C (32°F - 104°F).
*
NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.