7
Cd4
6
DIP STRIP
About 4 seconds after the meter auto-zeroes, “20” will appear on the display. Immediately dip the
eXact
®
Strip Micro Cd (DPD-1), Part No. 486633
into the
CELL
. During this
“20” SECOND
countdown timer, move the strip in a gentle back and forth motion (approx 2 strokes/sec).
Remove and discard the strip after “1” on the display disappears*
. The meter will begin
counting up for 100 seconds, at the end of which, the cursor will move across the display while
the meter prepares to measure the sample. Record result displayed (this result is automatically
stored in Cd4). After testing is completed, rinse cell immediately.
5
Dip the
eXact Strip Micro Glycine, Part No. 484014
into the
CELL
and immediately press
READ
.
This starts the
20 SECOND
countdown timer. During this time move the strip in a gentle back and
forth motion (approx 2 strokes/sec).
Remove and discard the strip after “1” on the display
disappears
. The meter will auto-zero. Get ready to dip
eXact
®
Strip Micro Cd (DPD-1)
strip.
DIP STRIP AND PRESS “READ”
4
Rinse the
CELL
at least 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\ZLWK
the water sample.
FILL METER WITH SAMPLE
3
Press and re-press the
MENU
button until the display shows the parameter
Cd4
.
SELECT TEST: Cd4
2
Press the
ZERO/ON
button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
1
Remove one (1)
eXact Strip Micro Glycine, Part No. 484014
and one (1)
eXact
®
Strip Micro
Cd (DPD-1), Part No. 486633
from the bottle before beginning the test. Set the strips in a dry,
convenient place and recap the bottle immediately.
REMOVE STRIP
Chlorine Dioxide Test Procedure
1
Remove one (1)
eXact
®
Strip Micro O
3
(DPD-4), Part No. 486634
from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
6
DIP STRIP AND PRESS “READ”
Dip the
eXact
®
Strip Micro O
3
(DPD-4), Part No. 486634
into the
CELL
and immediately press
READ
. This starts the
20 SECOND
countdown timer. During this time move the strip in a
gentle back and forth motion (approx 2 strokes/sec).
Remove and discard the strip after “1”
on the display disappears*
. The cursor will move across the display while the meter prepares
to measure the sample. Record result displayed (this result is automatically stored in O3). After
testing is completed, rinse cell immediately.
5
Press the
ZERO/ON
button. The cursor will move across the display, followed by
0.00 PPM
.
The sample is ready for testing.
ZERO METER*
2
Press the
ZERO/ON
button to power the meter on; the display will show all annunciators, then
the current MENU selection, followed by the last reading.
TURN METER ON
3
Press and re-press the
MENU
button until the display shows the parameter
O3
.
SELECT TEST: O3
4
Rinse the
CELL
at least 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\ZLWK
the water sample.
FILL METER WITH SAMPLE
Ozone Test Procedure
o
3
O3
*
NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.