45
11.0 A pp en dix
11.0 Appendix
SCU control unit
Technical data
Benefits
• Compact 6-channel actuator control
unit
• Single fault safety
• Overload and over-temperature
protection
• Approved for medical applications
• Easy to clean
• Low standby current
• Remote control RS232
Unit
SCU 1
SCU 5
SCU 9
Motor ports (DIN8)
#
6
3 or 6
3 or 6
Operating device ports (HD15)
#
3
3
3
Battery ports (DSub9)
#
1
1
1
Limit switch ports (HD15)
#
4
4
4
Single fault safety
yes/no
yes
yes
yes
Encoder processing
yes/no
yes
yes
yes
Input voltage
V
24 DC
120 AC
230 AC
Frequency
Hz
N/A
60
50
Input current (max)
A
30
6,5
3,3
Standby power
W
0,8
4,3
4,3
Output voltage
V DC
24
24
24
Output current (max)
A
30
18
18
Duty cycle: intermittent
min.
1 min./9 min.
1 min./9 min.
1 min./9 min.
Duty cycle: short time
min.
2
2
2
Ambient temperature
°C
+5 to +40
+5 to +40
+5 to +40
Humidity
%
5 to 85
5 to 85
5 to 85
Type of protection
IP
×4
×4
×4
Approvals
EN/UL
EN 60601–1
EN 60601–1
EN 60601–1
UL 60601–1
UL 60601–1
UL 60601–1
Weight without battery
kg
1,2
3,8
3,8
Weight with battery
kg
4,2
6,8
6,8