58
WT.050.590.000.DE.IM.0719
Functions
SFC
4.
Membrane sensor for chlorine
dioxide CD7
The CD7 membrane sensor enables the chlorine dioxide content
in all types of water to be determined selectively, also in high-purity
water (LF > 1µS/cm), without cross-sensitivity to chlorine, bromine
and hydrogen peroxide, but with cross-sensitivity to ozone and
peracetic acid. The sensor demonstrates no pH value dependency
(see „Technical data“) and is therefore very suitable for water with
various pH values.
Commissioning
Caution!
The electrode fingers (D) and membranes (G) are extremely
sensitive! Do not touch, soil or damage! Before unscrewing the
filled membrane cap (E), push the elastomer seal (H) to one side
to permit the inflow of air through the vent (F) underneath it (at the
cap label)! Otherwise, the membrane (G) may be damaged due to
the development of underpressure. Do not remove the light
yellowish-gray deposit of the reference electrode (K) or wipe it in
the direction of the gold working electrode (J)!
Flush eyes and skin immediately with water after contact. Rinse
away spilled electrolyte with water.
1
Screw off the membrane cap (E) from the electrode shaft (B)
and fill with the included gel electrolyte up to the top brim.
2
Rub gold working electrode (J) with the included lapping paper
(special emery). To do this, lay the lapping paper on a paper
towel, take hold of a corner, and using the tip of the vertically-
held electrode shaft (B), slide it once or twice over the rough
side of the lapping paper.
3
Check whether the elastomer seal (H) completely closes the
valve opening (F).
4
Then screw the filled membrane cap (E) slowly (by hand) back
onto the the electrode shaft (B).
Please note
As the excess electrolyte escapes through the valve opening (F)
under the elastomer seal (H), do not clamp it shut and do not press
it onto the elastomer seal (H).
A
D
E
G
H
J
K
B
F
I
C
W3T164493
Summary of Contents for WALLACE & TIERNAN SFC
Page 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Page 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Page 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Page 121: ...SFC Installation 5 121 Top hat rail assembly...
Page 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 123: ...SFC Installation 5 123 Top hat rail assembly...
Page 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Page 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Page 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Page 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Page 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Page 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Page 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Page 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Page 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...